Taiwan Biophotonic Corporation
85
Table 9. Electromagnetic Immunity
This oCare
TM
Pro 100 is intended for use in the
electromagnetic environment specified below. The customer
and/or user of the oCare
TM
Pro 100 should ensure that it is used
in such an environment.
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment –
Guidance
Electrostatic
Discharge
(ESD)
IEC 61000-4-2
±8 kV
contact
±15 kV air
±8 kV
contact
±15 kV air
Floors should be
wood, concrete,
or ceramic tile. If
floors are covered
with synthetic
material, relative
humidity should
be at least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for
power
supply lines
±1 kV for
input/output
lines
N/A
Mains power
quality should be
that of a typical
commercial or
hospital
environment.
Surge
IEC 61000-4-5
±1 kV
differential
mode
±2 kV
common
N/A
Mains power
quality should be
that of a typical
commercial or
hospital
Содержание oCare Pro100
Страница 1: ...Wrist worn Pulse Oximeter oCareTM Pro 100 Instructions for Use Ver P01A01 4...
Страница 14: ...oCareTM Pro 100 14 NCC...
Страница 35: ...Taiwan Biophotonic Corporation 35 Figure 3 Battery Charge...
Страница 38: ...oCareTM Pro 100 38 Figure 5 Adjustment of Watchband Length...
Страница 87: ...Taiwan Biophotonic Corporation 87 NOTE UT is the AC mains voltage before application of the test level...
Страница 92: ...oCareTM Pro 100 92 Note...
Страница 93: ...Taiwan Biophotonic Corporation 93 Note...
Страница 94: ...oCareTM Pro 100 94 Note...
Страница 95: ...Taiwan Biophotonic Corporation 95 Note...