![tBPC oCare Pro100 Скачать руководство пользователя страница 78](http://html1.mh-extra.com/html/tbpc/ocare-pro100/ocare-pro100_instructions-for-use-manual_3806528078.webp)
oCare
TM
Pro 100
78
Displayed Values
SpO
2
Update the last 8-beat moving average of the functional oxygen
saturation at one-second interval.
Pulse Rate
Update the last 8-beat moving average of the pulse rates at
one–second interval.
Principles of Operation
oCare
TM
Pro 100 uses a non-invasive pulse oximetry method to
measure the functional oxygen saturation in the blood. More
specifically, it measures the amount of oxygenated hemoglobin
as percentage of functional hemoglobin, including oxygenated
hemoglobin (HbO
2
) and deoxygenated hemoglobin (Hb).
Blood vessels, tissue and bone normally absorb a constant
amount of light over time. The arteriolar bed normally
pulsates and absorbs variable amounts of light during the
pulsations. The ratio of light absorbed is translated into
measurement of functional oxygen saturation (SpO
2
).
Содержание oCare Pro100
Страница 1: ...Wrist worn Pulse Oximeter oCareTM Pro 100 Instructions for Use Ver P01A01 4...
Страница 14: ...oCareTM Pro 100 14 NCC...
Страница 35: ...Taiwan Biophotonic Corporation 35 Figure 3 Battery Charge...
Страница 38: ...oCareTM Pro 100 38 Figure 5 Adjustment of Watchband Length...
Страница 87: ...Taiwan Biophotonic Corporation 87 NOTE UT is the AC mains voltage before application of the test level...
Страница 92: ...oCareTM Pro 100 92 Note...
Страница 93: ...Taiwan Biophotonic Corporation 93 Note...
Страница 94: ...oCareTM Pro 100 94 Note...
Страница 95: ...Taiwan Biophotonic Corporation 95 Note...