oCare
TM
Pro 100
10
10.
This device is designed to determine the percentage of
arterial oxygen saturation of functional hemoglobin.
Factors that may degrade pulse oximeter performance or
affect the accuracy of the measurement include the
following:
excessive ambient light
excessive motion
electrosurgical interference
blood flow restrictors (arterial catheters, blood
pressure cuffs, infusion lines, etc.)
anemia or low hemoglobin concentrations
improperly applied this device
poor pulse quality
cardiogreen and other intravascular dyes
carboxyhemoglobin
methemoglobin
dysfunctional hemoglobin
venous pulsations
11.
Do not perform any testing or maintenance on this device
while it is being used to monitor a user.
12.
This device is a precision electronic instrument and must
be repaired by qualified technical professionals. Field
repair of this device is not possible. Do not attempt to
open the case or repair the electronics. Opening the case
Содержание oCare Pro100
Страница 1: ...Wrist worn Pulse Oximeter oCareTM Pro 100 Instructions for Use Ver P01A01 4...
Страница 14: ...oCareTM Pro 100 14 NCC...
Страница 35: ...Taiwan Biophotonic Corporation 35 Figure 3 Battery Charge...
Страница 38: ...oCareTM Pro 100 38 Figure 5 Adjustment of Watchband Length...
Страница 87: ...Taiwan Biophotonic Corporation 87 NOTE UT is the AC mains voltage before application of the test level...
Страница 92: ...oCareTM Pro 100 92 Note...
Страница 93: ...Taiwan Biophotonic Corporation 93 Note...
Страница 94: ...oCareTM Pro 100 94 Note...
Страница 95: ...Taiwan Biophotonic Corporation 95 Note...