SPECIFICATIONS

XL-45HP

XL-50HP

XL-55HP

XL-50VHP

Dimensions

Diameter

20 in. (508 mm)

20 in. (508 mm)

20 in. (508 mm)

20 in. (508 mm)

Height

61 3/8 in. (1559 mm)

64 5/8 in. (1641 mm)

69 7/16 in. (1764 mm) 64 3/4 in. (1645 mm)

Weight

Empty (Nominal)

272 lb. (123 kg)

284 lb. (129 kg)

287 lb. (130 kg)

310 lb. (141 kg)

Capacity, Gross

176 liters

188 liters

208 liters

188 liters

Capacity, Useable Liquid

165 liters

176 liters

198 liters

176 liters

Weight of Contents Maximum

Based on DOT Rated Service Pressure

Carbon Dioxide

387 lb. (176 kg)

414 lb. (188 kg)

458 (208 kg)

381 (173 kg)

Oxygen

360 lb. (163 lb.)

385 lb. (175 kg)

426 (193 kg)

364 lb. (165 kg)

Nitrogen

252 lb. (114 kg)

269 lb. (122 kg)

298 lb. (135 kg)

240 lb. (109 kg)

Argon

438 lb. (199 kg)

467 lb. (212 kg)

518 lb. (235 kg)

443 lb. (201 kg)

Nitrous Oxide

368 lb. (167 kg)

393 lb. (178 kg)

435 lb. (197 kg)

N/A

Normal Evaporation Rate*

(% Capacity per Day)

Carbon Dioxide

0.75%0.75%

0.75%

0.80%

Oxygen

1.4%

1.2%

1.2%

1.5%

Nitrogen

2.2%

2.0%

1.9%

2.2%

Argon

1.4%

1.2%

1.2%

1.5%

Nitrous Oxide

0.75%

0.75%

0.75%

N/A

Gas Flow Rate @ NTP (STP)**

Carbon Dioxide

150 cfh (3.9 cu.m/h)

150 cfh (3.9 cu.m/h)

150 cfh (3.9 cu.m/h)

150 cfh (3.9 cu.m/h)

Oxygen, Nitrogen, Argon

350 cfh (9.2 cu.m/h)

350 cfh (9.2 cu.m/h)

350 cfh (9.2 cu.m/h)

350 cfh (9.2 cu.m/h)

Nitrous Oxide

110 cfh (2.9 cu.m/h)

110 cfh (2.9 cu.m/h)

110 cfh (2.9 cu.m/h)

N/A

Relief Valve Setting

350 psig

350 psig

350 psig

500 psig

(24 bar/2413 kPa)

(24 bar/2413 kPa)

(24 bar/2413 kPa)

(34 bar/3447 kPa)

Inner Container Bursting Disc

525 psig

525 psig

525 psig

750 psig

(36 bar/3620 kPa)

(36 bar/3620 kPa)

(36 bar/3620 kPa)

(52 bar/5171 kPa)

Dual Pressure Building/

Economizer Regulator***

Pressure Building Setting

300 psig

300 psig

300 psig

400 psig

(20.7 bar/2068 kPa)

(20.7 bar/2068 kPa)

(20.7 bar/2068 kPa)

(28 bar/2758 kPa)

Economizer Setting

320 psig

320 psig

320 psig

420 psig

(22 bar/2206 kPa)

(22 bar/2206 kPa)

(22 bar/2206 kPa)

(29 bar/2896 kPa)

Design Specifications

TC

4LM

4LM

N/A

N/A

DOT

4LM

4LM

4LM

4LM

Gaseous Capacity

Based on DOT Rated Service Pressure

@ NTP (STP)

Carbon Dioxide

3383 cu.ft. (89 cu.m)

3619 cu.ft. (95cu.m)

4003 cu.ft. (108 cu.m)

3330 cu.ft. (88 cu.m)

Oxygen

4350 cu.ft. (114 cu.m)

4651 cu.ft. (122 cu.m)

5146 cu.ft. (135 cu.m) 4397 cu.ft. (116 cu.m)

Nitrogen

3478 cu.ft. (91 cu.m)

3712 cu.ft. (98 cu.m)

4112 cu.ft. (108 cu.m)

3312 cu.ft. (87 cu.m)

Argon

4236 cu.ft. (111 cu.m)

4516 cu.ft. (119 cu.m)

5012 cu.ft. (132 cu.m) 4285 cu.ft. (113 cu.m)

Nitrous Oxide

3211 cu.ft. (84 cu.m)

3429 cu.ft. (90 cu.m)

3796 cu.ft. (106 cu.m)

N/A

Specifications are subject to change without notice.

* Vented N.E.R. based on Useable Liquid Capacity.

** Container pressure at or above factory Dual Pressure Building/Economizer Regulator setting.

*** Regulator has a pressure delta of 20 psig (1.4 bar/138 kPa).

3

Содержание XL-45HP

Страница 1: ...gulator Do not attempt to use or maintain this unit until you read and understand these instructions Do not permit untrained persons to use or maintain this unit If you do not fully understand these i...

Страница 2: ...if used in oxygen or nitrous oxide service although they may be acceptable for use with other cryogenic liq uids Use only Taylor Wharton recommended spare parts and be certain parts used on oxygen or...

Страница 3: ...cfh 2 9 cu m h 110 cfh 2 9 cu m h 110 cfh 2 9 cu m h N A ReliefValveSetting 350 psig 350 psig 350 psig 500 psig 24 bar 2413 kPa 24 bar 2413 kPa 24 bar 2413 kPa 34 bar 3447 kPa Inner Container Bursting...

Страница 4: ...tion q When loading or unloading the container from a truck use a hand truck lift gate crane or parallel loading dock Never attempt to manually lift the unit q To move the container over rough surface...

Страница 5: ...a mean of introducing heat from outside the container s insulated jacket to vaporize liquid as gaseous product is withdrawn The capacity of this circuit is sufficient to vaporize at flow rates up to...

Страница 6: ...re quired CGA connection that matches the gas service for which the container is configured The LIQUID Valve Liquid product is added or withdrawn from the container through the connection controlled b...

Страница 7: ...icator band This gauge is an indication of approximate container contents only and should not be used for filling liquid cylinders should be filled by weight Relief Devices These cylinders have a gas...

Страница 8: ...capacity is required an ac cessory external vaporizer is available When an external vaporizer is sued it must be connected to the USE valve and the regulator moved to the output of the external vapor...

Страница 9: ...ainer is being filled with CO2 it may be necessary to pressurize the container with gaseous CO2 before beginning the fill Pressure Transfer Filling Method Filling a liquid cylinder using the pressure...

Страница 10: ...to the VENT valve on the cylinder Both the fill line and the container should be pre cooled prior to beginning the fill process Proper full weight is determined by the previ ously explained method 1 O...

Страница 11: ...ervice Hose Length Connections s Fittings Number Inert N2 Ar Service 4ft 1 2m StainlessSteel LIQUID or VENT Valve CGA 295 to 3 8 in NPT 1700 9C65 6ft 1 8m StainlessSteel LIQUID or VENT Valve CGA 295 t...

Страница 12: ...and VENT valves then changing the liquid level gauge scale by changing its plastic cover and revising product decals Parts are available in kit form for each gas service as illustrated in the followi...

Страница 13: ...the change REGULATOR ADJUSTMENT RANGES Part No Normal Setting Range Delta 8816 1060 400 psig 300 to 600 psig 28 bar 20 7 to 41 bar 2758 kPa 2068 to 4137 kPa 6999 9018 300 psig 200 to 350 psig 20 psig...

Страница 14: ...roceeding 2 Close the on off valve and the Dump valve 3 Slightly open the high pressure cylinder valve 4 Set the high pressure regulator above the desired set point for the Pressure Building setpoint...

Страница 15: ...ill not hold liquid NER Testing If a loss of vacuum integrity is suspected the container s Normal Evaporation Rate NER should be checked The test measures the actual product lost over time so you can...

Страница 16: ...touch the side or bottom of the tube The yellow level indicator of the gauge should indicate a full level reading with the oxygen scale If the gauge fails to indicate a full liquid level the assembly...

Страница 17: ...drop the float rod straight through the guide ring inside the cylinder 4 To confirm that the rod is correctly positioned in the cylinder stop where you can still grasp the top of the rod see illustra...

Страница 18: ...to about 20 ft lbf 2 8 kgf m Leak check the connection of gauge body to the flange CAUTION When installing the gauge assembly care must be taken to ensure that the float rod is inserted through the gu...

Страница 19: ...alve Disassembly Instructions 1 Open valve by turning handwheel counterclockwise as far as it will go to release any trapped gas in the system 2 Using a screwdriver remove Handwheel Screw and Washer b...

Страница 20: ...nnet to 80 ft lbf 11 kgf m torque 6 Install Bonnet Washers over Stem on Bonnet 7 Place Handwheel over Stem on Bonnet 8 Install Seal 6 over Stem into recess of Handwheel 9 Install Seal Washer 5 over Se...

Страница 21: ...on container Use a rubber hammer to drive the rubber shock ring back into place 5 Using a in drill but drill holes through the rubber so that the carriage bolt slides in smoothly 6 The holes in foot r...

Страница 22: ...ce bursting bursting disc disc Pressurize container and check relief valve operation 5 3 Leaks in valves or 3 Leak test and repair leaks plumbing For valve repairs see Maintenance section 4 Cold liqui...

Страница 23: ...e Economizer Regulator set regulator when in use above relief valve setting Economizer side of regulator clogged or stuck open Container vents after This may be caused by Symptom should go away fill b...

Страница 24: ...re AL36590 0568 1 800 898 2657 in USA and Canada Accessories available for use with Taylor Wharton XL Series containers are Manifolds Automatic and Manual Container Hand Trucks Vaporizers adding up to...

Страница 25: ...06 9C14 2 Each 7 Safety Head 525 psig 36 bar 3620 kPa 1705 9C12 2 Each Safety Head 750 psig 52 bar 5171 kPa for XL 50VHP Only 7815 3085 2 Each 8 Relief Valve 500 psig 34 bar 3447 kPa not for C02 for X...

Страница 26: ...oxide 7355 4712 5 Each Decal Carbon Dioxide GL55 9C54 A R Decal Nitrogen GL55 9C51 A R Decal Oxygen GL55 9C52 A R Decal Argon GL55 9C53 A R Decal Nitrous Oxide GL55 9C55 A R Decal Warning 1700 9C07 4...

Отзывы: