![TASLER LTT500 Скачать руководство пользователя страница 32](http://html.mh-extra.com/html/tasler/ltt500/ltt500_manual_821036032.webp)
5. Channel setup
Operation Manual Version 1.6
32
5.2.1. OFF
No input circuit for this channel is selected. No input signal will be transferred to the output.
5.2.2. VOLT
The Voltage input signal will be transferred to the output. The amplification or attenuation depends on the
selected input range in the next column. With the range ±5V (1 : 1, or 0dB) the output gives the input signal 1 :
1 out. Following input ranges can be selected:
As next you select in the column Coupling the input coupling circuit for this channel. You can decide between
following configurations:
Coupling
ODU-Pin
Description
Single Ended DC+
Pin 3
Pin 8
Input +
Signal Ground
Single Ended DC-
Pin 2
Pin 8
Input -
Signal Ground
Differential Ended DC
Differential Ended 3Hz High Pass Filter
Pin 3
Pin 2
Input +
Input -
Single Ended 3Hz High Pass Filter
Single Ended 16Hz High Pass Filter
Pin 3
Pin 8
Input +
Signal Ground