DESCRIPTION OF EQUIPMENT 

1.1 

Electrical Specification 

1.1.1 

Supply Requirements 

 

KV5-100 mk3  115/230V 

10% 600VA max 

1.1.2 

Output Specifications 

 

The output voltage, current and duty cycle are as follows: 

Unit 

Output voltage 

Output current 

5 min on/ 

15 min off 

1 min on/ 

15 min off 

40 seconds on/ 

15 min off 

KV5-100 mk3 

0-5kV 

100mA 

 

 

One  side  of  the  high  voltage  transformer  output  is  connected  to  the  earth  system  of  the 
equipment. 

1.1.3 

Zero Voltage Interlock 

 

The equipment is fitted with a zero volt interlock system on the output control.  This interlock 
prevents  the  output  being  energised  unless  the  output  control  is  in  the  minimum  output 
position. 

1.1.4 

External Interlock 

 

The  equipment  is  fitted  with  a  external  interlock  system.      The  interlock  socket  on  the  back 
panel must have a short circuit across it terminals to enable the output to be energised. 

 

Note:  the  open  circuit  voltage  across  the  interlock  terminals  is  240Vac  and  is  connected 
directly to the mains supply. 

 

The  unit  is  supplied  with  a  plug  fitted  to  the  external  interlock  socket  on  the  rear  of  the  unit.  
The  plug  contains  a  shorting  link  to  enable  the  unit.    If  external  interlocks  are  used,  the 
interlocks  should  replace  the  shorting  link.    Please  note  that  the  output  will  not  switch  on 
without either the link fitted in the plug or a closed external interlock circuit. 

1.2 

Output Voltage Control 

 

The  output  from  the  equipment  is  controlled  from  zero  by  means  of  the  output  control.    To 
increase the output voltage turn the control knob in a clockwise direction. 

Note:

 

The  output  cannot  be  energised  unless  the  regulator  knob  is  set  to  zero,  thus  operating  the 
zero voltage interlock switch. 

Содержание KV5-100 mk3

Страница 1: ...igh Voltage AC Test Set Type KV5 100 mk3 DESIGNED AND MANUFACTURED BY T R Test Equipment Limited 15 16 Woodbridge Meadows Guildford Surrey GU1 1BJ United Kingdom Telephone 01483 207428 e mail sales tr...

Страница 2: ...1...

Страница 3: ...is provided on the unit for connection to a local earth if necessary Ensure that the unit is properly grounded before making any connections to inputs or outputs Output ratings must be observed to pr...

Страница 4: ...d to both the test set and the test specimen As a minimum the current capacity of the ground leads should exceed that necessary to carry the maximum possible ground current The effect of ground potent...

Страница 5: ...The following safety symbols appear in this manual CAUTION This action or procedure may be dangerous if not carried out correctly and may cause damage to the equipment or connected equipment WARNING T...

Страница 6: ...tion 8 1 3 1 Variable overload 8 1 4 Metering 9 1 5 Fault Burn Facility 10 1 5 1 Fault Burning 10 1 5 2 Burn Switch 10 1 6 High Voltage Connection 10 1 7 Construction 10 2 OPERATION 11 2 1 Safety 11 2...

Страница 7: ...6 3 MAINTENANCE 16 4 STANDARD ACCESSORIES 17 5 OVERALL PERFORMANCE SPECIFICATION 18 6 REVISION 19...

Страница 8: ...system The interlock socket on the back panel must have a short circuit across it terminals to enable the output to be energised Note the open circuit voltage across the interlock terminals is 240Vac...

Страница 9: ...ip levels are adjustable in two ranges each of 6 steps giving 2 12 and 20 120 rated current Unit Trip current ranges X1 X10 KV5 100 mk3 2 4 6 8 10 12mA 20 40 60 80 100 120mA The range change and selec...

Страница 10: ...th the following metering ranges using analogue moving coil meters Unit Metering Ranges Voltage Current Low High X1 X10 KV5 100 mk3 3kV 6kV 10mA 100mA The ranges on both instruments are selected by me...

Страница 11: ...lower output voltages the burn current limit reduces in proportion to the output voltage When the burn switch is in the out position the maximum current is approximately 2A for short periods of time...

Страница 12: ...supply or HV unit interlock to ensure that the unit cannot be switched on unless the area is secure Refer to BS EN50191 2001 and IEEE Std 510 1983 for further details of suitable test enclosures The t...

Страница 13: ...procedures must be implemented to accepted standards 2 1 2 1 Interlocks The KV5 100 mk3 series is provided with an external interlock circuit that may be used to link to interlock switches on the high...

Страница 14: ...itch Selects kV meter range H Earth output lead Earth output to object under test J HV Output ON pushbutton Switches HV output on Illuminated red when output is on K Variable trip level select Selects...

Страница 15: ...of Operation 2 4 1 Connections WARNING Before making any connections please ensure that you are aware of all hazards relating to the system and environment in which it is operating Ensure that the ke...

Страница 16: ...p button is released and that the regulator is in the fully anti clockwise position Press the HV ON pushbutton The HV output will now energise and the red HV on indicator in the pushbutton will light...

Страница 17: ...t and on the output leads Ensure that the unit is disconnected from the mains before removing any covers It is advisable to check the carbon brush on the variable transformer for signs of wear on a re...

Страница 18: ...ANDARD ACCESSORIES The following items are provided with the equipment a Mains input lead b 2 keys for mains ON OFF switch c Operating Maintenance Manual d 1x F5A 1 inch fuse and 1x T4A 20mm fuse KV5...

Страница 19: ...stance at 500V DC 10 megohms between mains input and frame Applied voltage test 2 0kV RMS for 1 minute between mains input and frame Accuracy of instruments kV meter 1 5 of FSD mA meter 2 5 of FSD Ins...

Страница 20: ...19 6 REVISION Product Type KV5 100 mk3 series High Voltage AC Test Systems File KV5 100mk3 series manual v1 Author P Cole Issue Date 1 18 09 20 Modified By M Clancy Checked By G Bond Date 06 10 20...

Страница 21: ...20 Circuit diagram A2 002174 KV5 100 mk3...