Communication Functions
51
Registering and Unregistering Remote Units for
"Visitor Entry"
[
Visitor Entry]
This function allows for non-regular wireless communication between
an RTR-500DC and Remote Units which have been registered in
RTR-500 Series Base Units.
Up to 42 Remote Units can be assigned “Visitor Entry” status.
1.
In the Main Menu, open [WL Visitor Entry].
2.
To assign “Visitor Entry” status to a Remote Unit, place the target
Remote Unit face down on the RTR-500DC connected to the PC,
and select [Accept Visitor RU].
3.
A [Completed] message will appear with the Serial No. and Unit
Name* of the Remote Unit which has been assigned “Visitor Entry”
status.
* This shows the Remote Unit
Name entered when Remote
Unit Registration was carried
out.
4.
To release “Visitor Entry” status, select [Release One] and choose
Registration Error Messages
[Unregistered]
This message appears when communication is carried out
between the RTR-500DC and a Data Logger which is in its
original factory default settings or has been re-initialized by
using the RTR-500DC Settings Utility.
[Duplication] / Remote Unit Name (alternating display)
They will be alternately displayed when communication is
carried out with any Remote Unit which has been already
registered for “Visitor Entry” or any Remote Unit which has the
same Remote Unit Number or the same Group Registration
Info.
[Incompatible]
This message appears when communication is carried out
between the RTR-500DC and an incompatible Data Logger.