Press
Right
key twice to blink the
of
under the
GV
.
(9)
Pressing
Up
key will open the solenoid valve and start gas flow.
(10)
Use a soft brush to apply soapy water to each of the gas line joints and check for bubbles.
Bubbles appearing at any of the gas line joints indicate the presence of a gas leak. Cease
using the APM-7 and contact the manufacturer or an authorized distributor for repairs.
(11)
After completion of the inspection, press the
RESET
key. This will shut the solenoid valve.
(12)
Turn OFF the main switch and remove the power plug from the wall socket.
(13)
Remove signal connecting cable and power connecting cable from the test unit.
(14)
Attach side covers and then rear cover to the test unit.
(15)
Re-connect
signal
connecting
cable and power connecting cable to the test unit.
(16)
Reinsert the power plug into the wall socket to conclude gas leak inspection.