37
GE
SW
2
Fig. 31 MMB Board
(12) Turn the MAIN switch ON, and confirm the character of “Electric” appears on the Main
screen. Gas solenoid valve will never open by this operation.
(13) Change the filament output level to the value of 75 to 90%. See page 9 Chapter
2
.
1
.
(14) Adjust the filament output level and dipping angle again so that the obtained flash
point will correct. Be careful that the ignition coil does not touch the test cover.
Set Screw for
Dipping Angle
Fig. 32 Dipping Angle