10
electric ignition configuration, enter provisional output value or 10% lower than the present setting.
Table 1 Filament Output Values
Provisional Output Value
Actual AC
Source
Gas Ignition
Electric Ignition
100V 30% 65%
110V 25% 60%
115V 25% 55%
120V 20% 50%
220V 30% 65%
230V 25% 60%
240V 20% 55%
- Gas Ignition: After entering new output value, push START key to confirm the gas can be
ignited by filament within a minute. If the tester is not used for a long time, it may take a few
minutes to bleed the air out of the gas piping. If the gas is coming out but not ignited by
filament, increase output value.
- Electric Ignition: After entering new output value, run a test using CRMs. Increase set output
value if the obtained flash point was higher than expected; decrease it if lower than expected.
If the obtained flash point was within a range, decrease the set output value by 5%, and then
run a test again. The lowest output when the obtained flash point is still within the acceptable
range is the best setting for the extended life of the igniter.
Fig. 4 Filament Output Setting Screen
By selecting
Ignit
(
off
is blinking), and pressing
Up-Down
key to turn ignition
on
, electric igniter
will be lighting and actual output (A) can be seen.
Fig. 5 Filament Output Setting Screen (lighting)
APM Filament Ignit.source
Gas flame
Set output Ignit
40 % off
APM Filament Ignit. sourc
Gas flame
Set output Ignit Current
40 % on 3.8(A)