18
Sample programs:
(1)
SHORTTEST
STEP1:
The extraction head will mount spin tips on the left half part of a spin
tip rack. After mounting the spin tips, the user can replace the spin tip rack
with a reagent deep well plate. After placement of reagent deep well plate,
press “
Pause
” to continue.
STEP2:
The 9600 LH will perform mixing for 0.1 minutes and collecting for 10
seconds in plate 1. After that, it will move out the plate holder and beep
(Pause: ON setting). The user can place the next reagent deep well plate and
press “
Pause
” button to continue.
STEP3:
The 9600 LH will perform mixing for 0.1 minutes and collecting for 10
seconds in plate 2. After that, it will move out the plate holder and beep
(Pause: ON setting). The user can place the next reagent deep well plate and
press “
Pause
” button to continue.
STEP4:
The 9600 LH will perform mixing for 0.1 minutes and collecting for 5
seconds in plate 3. After that, it will move out the plate holder and beep
(Pause: ON setting). The user can place the next reagent deep well plate and
press “
Pause
” button to continue.
STEP5:
The 9600 LH will perform mixing for 0.1 minutes and collecting for 10
seconds in plate 4. After that, it will move out the plate holder and beep
(Pause: ON setting). The user can place the next reagent deep well plate and
press “
Pause
” button to continue.
STEP6:
The 9600 LH will perform mixing for 0.1 minutes and ejection of spin
tips in the left half part of reagent deep well plate.
(2)
EJECT-TIP
STEP1:
The extraction head will mount spin tips from the left half part of spin
tip rack. After mounting of spin tips, it will move out plate holder. The user can
take away the spin tip rack and place an empty deep well plate. After that,
press “
Pause
” button to continue.
STEP2:
The 9600 LH will eject spin tips in the left half part of deep well plate.
Содержание Maelstrom 9600 LH
Страница 1: ...TAN Bead Nucleic Acid Extractor Non Sterile Maelstrom 9600 LH User Manual...
Страница 7: ...7 Equipment list No Item 1 Power cord 2 A plate of spin tips 3 Magnetic separator 4 User Manual...
Страница 9: ...9 Side view of extraction platform Shield plate...
Страница 10: ...10 Regulators of the sensors for detecting falling spin tip Control panel Buttons on control panel...
Страница 12: ...12 Set the Temperature UV LAMP Confirm the Selected Program Mount the Spin Tips...
Страница 13: ...13 Eject the Spin Tips STATUS RUNNING STATUS PAUSE STATUS STOP STATUS END...
Страница 17: ...17 NOTE...
Страница 22: ...22 6 2 Communication between M9600 LH and Liquid Handling System...
Страница 23: ...23 6 4 Communication CRC Cyclical Redundancy Check...