7. Product Performance
Consistency
The consistency of extracted lysates on the same machine and different machines
(within-run precision) is the variability of the results from well to well. The coefficient
of variation of nucleic acid extraction concentration is less than 5%. It is in the range
of acceptable statistical analysis.
▓
Nucleic Acid Yield
N=96
M96LH00001
M96LH00002
Average
Yield
N=96
Standard
Deviation
(SD)
Coefficient
of Variation
(CV%)
Average Yield
(ng/
l)
30.23
41.99
36.11
0.95
2.52%
Standard
Deviation (SD)
0.565
1.336
Coefficient of
Variation
(CV%)
1.87%
3.18%
▓
Ratio (A260/A280)
N=96
M96LH00001
M96LH00002
Average
Yield
N=96
Standard
Deviation
(SD)
Coefficient
of Variation
(CV%)
Average Yield
(ng/
l)
1.89
1.88
1.89
0.008
0.39%
Standard
Deviation (SD)
0.007
0.008
Coefficient of
Variation
(CV%)
0.37%
0.41%
Содержание Maelstrom 9600 LH
Страница 1: ...TAN Bead Nucleic Acid Extractor Non Sterile Maelstrom 9600 LH User Manual...
Страница 7: ...7 Equipment list No Item 1 Power cord 2 A plate of spin tips 3 Magnetic separator 4 User Manual...
Страница 9: ...9 Side view of extraction platform Shield plate...
Страница 10: ...10 Regulators of the sensors for detecting falling spin tip Control panel Buttons on control panel...
Страница 12: ...12 Set the Temperature UV LAMP Confirm the Selected Program Mount the Spin Tips...
Страница 13: ...13 Eject the Spin Tips STATUS RUNNING STATUS PAUSE STATUS STOP STATUS END...
Страница 17: ...17 NOTE...
Страница 22: ...22 6 2 Communication between M9600 LH and Liquid Handling System...
Страница 23: ...23 6 4 Communication CRC Cyclical Redundancy Check...