7. Product Performance 

 
Consistency

 

The consistency of extracted lysates on the same machine and different machines 

(within-run precision) is the variability of the results from well to well. The coefficient 

of variation of nucleic acid extraction concentration is less than 5%. It is in the range 

of acceptable statistical analysis.  

 

 Nucleic Acid Yield 

N=96 

M96LH00001 

M96LH00002 

Average 

Yield 

N=96 

Standard 

Deviation 

(SD) 

Coefficient 

of Variation 

(CV%) 

Average Yield 

(ng/

l) 

30.23 

41.99 

36.11 

0.95 

2.52% 

Standard 

Deviation (SD) 

0.565 

1.336 

 

 

 

Coefficient of 

Variation 

(CV%) 

1.87% 

3.18% 

 

 

 

 

▓ 

Ratio (A260/A280) 

N=96 

M96LH00001 

M96LH00002 

Average 

Yield 

N=96 

Standard 

Deviation 

(SD) 

Coefficient 

of Variation 

(CV%) 

Average Yield 

(ng/

l) 

1.89 

1.88 

1.89 

0.008 

0.39% 

Standard 

Deviation (SD) 

0.007 

0.008 

 

 

 

Coefficient of 

Variation 

(CV%) 

0.37% 

0.41% 

 

 

 

 
 
 
 
 
 

Содержание Maelstrom 9600 LH

Страница 1: ...TAN Bead Nucleic Acid Extractor Non Sterile Maelstrom 9600 LH User Manual...

Страница 2: ...1 Introduction 4 2 Overview 8 3 Installation 11 4 Get started 11 5 Edit program 14 6 Communication 21 7 Product Performance 24 8 Cleaning and maintenance 26 9 Disposal 26 10 Patent 27 11 Appendix 27...

Страница 3: ...nying documents Hot surface risk of burns Watch your fingers and your hands The symbol indicates that the product should not be discarded as unsorted waste but must sent to separate collection facilit...

Страница 4: ...tating the spin tips will mix suspension Extended magnetic rods can collect magnetic beads efficiently Purified nucleic acids can be obtained after lysis nucleic acids absorption wash and elution Flow...

Страница 5: ...weight of Maelstrom 9600 LH 30 kg Do not use Maelstrom 9600 LH in a location where there is large temperature variability or high humidity Temperature 4 40 Maximum operating altitude 2000 m Storage an...

Страница 6: ...ubstances Users should adhere to their institutional guidelines for the handling and disposal of all infectious substances used with this device It is important to clean the device after every use If...

Страница 7: ...7 Equipment list No Item 1 Power cord 2 A plate of spin tips 3 Magnetic separator 4 User Manual...

Страница 8: ...ions L 54 2 x W 28 5 x H 26 8 cm Controller dimensions 32 x 30 x 14 cm Power supply 100 240 V Capacity 96 samples per run Process volume 50 l 1500 l Spin speed 500 3000 rpm Heater 1 heated plates Magn...

Страница 9: ...9 Side view of extraction platform Shield plate...

Страница 10: ...10 Regulators of the sensors for detecting falling spin tip Control panel Buttons on control panel...

Страница 11: ...d Turn on the main power switch located on the back of controller to the on position Once Maelstrom 9600LH is switched on the instrument will perform a self test and calibration After booting you will...

Страница 12: ...12 Set the Temperature UV LAMP Confirm the Selected Program Mount the Spin Tips...

Страница 13: ...13 Eject the Spin Tips STATUS RUNNING STATUS PAUSE STATUS STOP STATUS END...

Страница 14: ...ller via AutoMag Version 5 2 program 5 1 Connection 1 Switch on Maelstrom 9600 LH 2 Connect A of transmission cable to the Maelstrom 9600 LH controller Fig 2 3 Connect B to the USB port of PC Fig 3 Fi...

Страница 15: ...nter 0 as a default Password maximum characters 8 C COM port In drop down menu select the used USB port NOTE The used USB port can be checked in window system control device manage port D Controller v...

Страница 16: ...Eject Pre Heat Warm up temperature 0 150 Heating Execute program temperature 0 150 Mixing M Spin time 0 999 minutes Speed RPM Spin speed 500 3000 RPM Shuttle Spin at the lower and the upper alternatel...

Страница 17: ...17 NOTE...

Страница 18: ...STEP4 The 9600 LH will perform mixing for 0 1 minutes and collecting for 5 seconds in plate 3 After that it will move out the plate holder and beep Pause ON setting The user can place the next reagen...

Страница 19: ...selected cells Copy Copy data of a whole row Paste Paste copied data Move up Move up a whole row of data Move down Move down a whole row of data Insert Insert a row of blank space Delete You can delet...

Страница 20: ...rom the PC to the machine B A Example of program transmission import There are 3 programs installed in the controller You can use AutoMag to retrieve these programs Type 0 and press import button to g...

Страница 21: ...transfer the program from the PC to the controller Set 1 1 and press Export button to transfer program 1 to the controller Set 0 1 and press Export button to transfer program 0 and program 1 to the c...

Страница 22: ...22 6 2 Communication between M9600 LH and Liquid Handling System...

Страница 23: ...23 6 4 Communication CRC Cyclical Redundancy Check...

Страница 24: ...acceptable statistical analysis Nucleic Acid Yield N 96 M96LH00001 M96LH00002 Average Yield N 96 Standard Deviation SD Coefficient of Variation CV Average Yield ng l 30 23 41 99 36 11 0 95 2 52 Stand...

Страница 25: ...96 Cycle 1 Cycle 2 Cycle 3 Cycle 4 Average Yield N 128 Standard Deviation SD Coefficient of Variation CV Average Yield ng l 25 3 24 7 25 3 26 5 25 0 0 588 2 35 Standard Deviation SD 0 005 0 005 0 007...

Страница 26: ...magnetic rods wipe with a soft cloth dampened with 70 ethanol If the magnetic rods cannot be cleaned please contact TANBead Tel 886 3 3167568 for technical assistance 9 Disposal In case the product is...

Страница 27: ...anada CA2862946 Japan JP6151735 Korea 10 1696517 WIPO WO2016127292 China CN104971638B Taiwan I526245 11 Appendix Transport To avoid damaging the linear side through the transport please use EPE buffer...

Страница 28: ...28 Manufacturer Taiwan Advanced Nanotech Inc 4F No 188 Wenhe Rd Guishan Dist Taoyuan City 333 Taiwan R O C TEL 886 3 3167568 http www tanbead com en...

Отзывы: