INTRODUCTION TO THE TC SERIES
The TAMSON model TC baths are used for precise
temperature control. This series offers a choice in
relatively large bath volumes.
6.1
General
The TAMSON baths are constructed throughout from
corrosion-resistant
– stainless steel and brassTeflon –
materials. Outher casing is zinc plated powder coated
steel. The bath is effectively insulated against heat loss
by a layer of Armaflex® rubber between the inner tank
and outer casing.
6.2
Pump
A circulation pump is built-in to guarantee an uniform
temperature distribution within the bath and providing the
possibility to circulate through a closed external system.
The standard pump offers a motor protection (small
thermal fuse) which is placed on the back panel of the
motor compartment. This protects the motor from
excessive loads.
When pumping viscous liquids the thermal fuse can
be activated. The motor fuse can be reset by pressing
the button.
The inlet and outlet of the pump are fitted with 8 mm hose
connections and 1/4” threaded fittings for metal tubing.
The outlet is provided with a small disc, stopping the
circulation action.
Before the fluid can begin circulating through
external systems this disc must be removed by
unscrewing the fitting and hose connection
removing the disc.
6.3
Cooling
The TC-bath is equiped with a cooling coil. Circulation of
fluid, with a lower temperature than the bath fluid, through
this coil will cool down the bath temperature. To get stable
control of the bath when using the cooling coil, the
temperature of the cooling fluid has to be approximately
5°C lower than the set point temperature of the bath.
Cooling is need for set point temperatures of
approximately 42°C and lower.