7.3
Backside of the thermostatic bath
1
Cooling fan
2
RS 232 Communication
3
In and outlet pump
side view connection:
4
Mains cable
5
Motor fuse
6
Cooling coil, 1/4" outher thread
7
Switch for fan (only for TC16 with p/n 31T0671/
31T0861)
8
Fan (only for TC16 with p/n 31T0671/31T0861)
7.4
Selecting parameters
The front panel layout shows the following four operating
keys:
Press shortly and following is displayed
SP.1 - Temperature set point in °C,
tU.n
-
Tuning the bath (“Atune”),
Pb
- Changing the tuning (PID) parameters "P"value
ti
- Changing the tuning (PID) parameters "I"value
td
- Changing the tuning (PID) parameters "D"value
OFSt
- Change temperature readout offset (± 5°C)
Each time is pressed another option is displayed.
Up and Down keys allow changing the listed value. All
changed values, like set point and PID parameters, will
be kept in memory when pressing the page key to leave