TP-70BS (Rev 3.0)
- 33 -
TALON TECH CO. LTD.
6-8
Notice on Pump Operation
1. During Pump is under process (RUN OR CYCLE)
don’t
try to modify the data.
(Please do it after Pump had stop properly)
▶
In this term of
“
modify data
”
means changing recipe, cycle, CAL value etc. If you modified the
data during pump operation, BUSY screen will be pop-up and it will be not saved.
▶
Screen moving in touch pad is OK.
▶
In case of Encompass type, during dispensing with key pad / Touch pad, the pump works
properly even if the warning alarm occurs. After the pump stops, the alarm is clear
automatically (Alarm Auto Clear). (During revising data, when the alarm occurs, the data
under modification don
’
t store but the previous data back-up.
2. Caution for pump setting
▶
Dispense Volume => set the volume range between 100(1cc) ~ 1000(10cc)
▶
Reason :For protect the pump , upper limit alarm set.
▶
Dispense Time =>It should be bigger than Dispense Volume.It should be smaller than system
recipe step time.
▶
Reason :As using the high cp PR , sudden pump operation could have the motor damaged and
could cause the MICRO BUBBLE.
▶
Reload Time =>depend on PR cp.
▶
Typically, high cp PR case, adjust 3~8 times reload time than dispense volume.
ex) Dispense Reload
Vol[
300
] Time[500] Vol[300] Time[
900
]
▶
This would be main reason of bubble issues. So please watch out for this.
3. Please do not press too hard to any of pump inner joint or others.
4. Please do not press too hard to Buffer Tank.
900
~ 2400