Using in Series/Parallel Operation
134
8. Press the FUNC. key of the Slave.
9. Turn the VOLTAGE dial of the Slave unit to
display the setting item number 38 in the Number
Display (voltage).
* At that time, turn off the Number Display (current).
10. Adjust the CURRENT dial of the Slave so that the
display of the Slave unit’s digital multimeter for
output voltage measurement is the same as the
display of the Master unit’s digital multimeter for
output voltage measurement.
11. Press the FUNC. key of the Slave unit to set the
full-scale calibration value of output voltage
setting.
12. This completes offset full-scale calibration for the
output voltage of the Slave unit in series
operation.
At the Master unit, set the constant voltage
setting value and the constant current setting
value to the original ones and start using.
Содержание HX-S-G2 Series
Страница 165: ...Special Loads 165 This page is blank...
Страница 169: ...Maintenance 169 This page is blank...
Страница 211: ...Specifications 211 External Diagrams 1 HX S 030 200G2 F I...
Страница 212: ...Specifications 212 2 HX S 030 400G2 F I...
Страница 213: ...Specifications 213 3 HX S 060 100G2 F I...
Страница 214: ...Specifications 214 4 HX S 060 200G2 F I...
Страница 215: ...Specifications 215 5 HX S 0120 50G2 F I...
Страница 216: ...Specifications 216 6 HX S 0120 100G2 F I...
Страница 217: ...Specifications 217 7 HX S 0400 15G2 F I...
Страница 218: ...Specifications 218 8 HX S 0400 30G2 F I...
Страница 219: ...Specifications 219 9 HX S 0500 12G2 F I...
Страница 220: ...Specifications 220 10 HX S 0500 24G2 F I...
Страница 221: ...Specifications 221 11 HX S 01000 6G2 F I...
Страница 222: ...Specifications 222 12 HX S 01000 12G2 F I...
Страница 223: ...Specifications 223...
Страница 239: ......