If any of the problems described below occur, conduct the checks outlined before requesting repair.
Problem
Check Points
Recovery Procedures
Chair does not move.
Overheating Prevention Device
Daily Inspection
Before Requesting Repair
19
Is the power
turned on?
These are not malfunctions
Is the power plug correctly
inserted?
Is the power switch turned on?
Has a fuse of the shop's
switchboard or breaker been
tripped?
Has the chair been continuously
operated for an extended period?
Correctly insert the power plug.
(p. 10)
Turn on the power. (p. 10)
Replace the fuse of the switchboard
or turn on the breaker.
The problem is due to the operation
of the overheating prevention device.
⇒
Leave the chair for approximately
10 min. to allow it to return to the
normal condition.
This is a protective circuit that prevents
abnormal overheating of internal
mechanisms due to continuous
operation and similar causes.
The chair might stop moving
temporarily.
Is the
switchboard
operating
normally?
Is the power
plug correctly
inserted?
Abnormal sound at startup
Slow movement of the chair
Sound when rotating the chair
Such sound is due to the internal hydraulic fluid being cold. It is not
a functional fault and should stop shortly thereafter.
Such movement is due to the internal operating fluid being cold.
Raise the room temperature or continue operating for a short
period. It will recover to the normal condition.
This is the sound of the steel balls rolling when the chair is rotated.
The level of this sound varies depending on the environment where
the chair is installed.
If any of the problems described above are inspected and the chair does not work normally even after
performing the recovery procedures, immediately stop using the chair, turn off the power, unplug the power
plug, and contact your closest dealer or Takara Belmont.
Perform inspection according to the "Daily Inspection Procedures" indicated on pages 20 and 21 of this
operation manual in order to ensure safe use of this chair.
Perform daily inspection according to the "Daily Inspection Procedures". Using the chair
without performing daily inspection can result in malfunction or accident.
WARNING