32
Customer Service Toll Free: 833.3TACTILE (833.382.2845)
Table 8: Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The Entre Plus system is intended for use in the electromagnetic environment specified
below . The customer or the user of the Entre Plus should assure that it is used in such an
environment .
Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment –
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 8 kV contact
+/-2 kV, +/-4 kV,
+/-8 kV, +/-15 kV air
Floors should be wood,
concrete or ceramic
tile . If floors are covered
in synthetic material,
the relative humidity
should be at least 30% .
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 100 kHZ
repetition
± 2 kV for power
supply lines
± 100 kHZ
repetition
Mains power quality
should be that of a
typical commercial or
hospital environment .
Surge
IEC 61000-4-5
± - .5 kV
± -1 kV line to line
± - .5 kV
± -1 kV
± -2 kV line to earth
± - .5 kV
± -1 kV line to line
± - .5 kV
± -1 kV
± -2 kV line to earth
Voltage dips, short
interruptions and
voltage variations
on power
adapter lines
IEC 61000-4-11
Voltage dips
0% UT; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270°
and 315° 0% UT;
1 cycle and 70% UT;
25 periods at 50Hz,
30 periods at 60Hz
Single phase: at 0°
Voltage
Interruptions 0%
UT; 250 periods at
50Hz, 300 periods
at 60Hz
Voltage dips
0% UT; 0,5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270°
and 315° 0% UT;
1 cycle and 70% UT;
25 periods at 50Hz,
30 periods at 60Hz
Single phase: at 0°
Voltage
Interruptions 0%
UT; 250 periods at
50Hz, 300 periods
at 60Hz
Mains power quality
should be that of a
typical commercial or
hospital environment .
If the user of the Entre
Plus system requires
continued operation
during power mains
interruptions, it is
recommended that
the device be powered
from an uninterruptible
power adapter .
NOTE U
T