background image

 

TAC-H

2

S/H

2

O/O

2

 

Pg. 4

TEXAS ANALYTICAL CONTROLS

  

4418 Bluebonnet Drive  |  Stafford, TX 77477  |  (281) 240-4160  |  

www.tac-controls.com

2  Installation

2.5.1    Optional Isolated 4-20 mA

An optional isolated 4-20 mA output is available, 

offering linear users definable current output that is 

proportional to the gas concentration. It typically is 

connected to a recorder or similar instrument, such as an analog input to a remote SCADA unit. 

The connections are labeled "+" and "

-

". Its power is provided internally/self-powered. You can 

purchase an isolated, loop-powered device (Part # FX-742) from Texas Analytical Controls.

2.4  Electrical Connections

The input power and the 4-20mA connections are in the analyzer’s enclosure. The power switch is in the 

upper left-hand corner of the enclosure. Use it to recycle the power as needed.

If ordered, Alarm 1 and Alarm 2 are single-pole, double-throw, form C (dry contact– requires external 

power). The contacts are rated for 2 amps at 125 volts AC and 4 amps at 30 volts DC. The customer 

connection board includes three terminals for each relay: common (C), normally open (NO) and normally 

closed (NC). These relays are fail-safe connected: When the analyzer is off or in an alarm condition, C and 

NC are connected. When the analyzer is on, and no alarms are activated, C and NO are connected.
Each analyzer comes standard with an orange, 2 pin connector for flow fail signal. A 5 AMP fuse is 

included to protect the analyzer from an overload. A spare fuse is also provided.
The analyzer draws approximately 1 amp, depending on its configuration. If the analyzer is powered 

by solar panels, the size of the solar panel and battery is extremely important. If the analyzer’s power 

requirements are not met, the controller may lose its program.

2.5  Communication Outputs

X7 Series analyzers come, standard, with a self-powered 4-20mA, Modbus via RTU serial port MJ1/2,  

and RS232/485 via LAN. The Ethernet connector is used for the RS 232/485. Consult tech support to 

change to the RTU port. 

to back 

of panel 

BOTTOM VIEW 

FRONT VIEW 

NO

C

NC

+ / -

+ / -

12v DC

+

G

4-20 mA OUT

+ / -

+ / -

POWER IN

L1 L2 G

RELAY

FLOW FAIL

+

+

+

+

OFF

ON

ETHERNET

5A

FUSE

SPARE

FUSE

ETHERNET

PORT

STANDARD

IN

SAMPLE

IN

SAMPLE

VENT

SENSOR

VENT 1

SENSOR

VENT 2

Power / Communication

DI

03/15/23

Texas Analytical Controls Inc.

SHEET 1 of 1

TV

REV - 1.0

PLEASE NOTE:

CONFIGURATION MAY BE 

DIFFERENT PENDING THE

ANALYZER ORDERED

*NOTE: Unused hubs 

should retain plastic 

cap(s) securely in place 

when not in use to keep 

analyzer internals free 

from bugs and dirt.

Remove Plastic 

Cap(s) As Needed 

Before Installation

*

Power the instrument with a source that is not 

likely to fail or to be accidentally turned off. The 

analyzer will begin operating as soon as the power 

is connected, drawing approximately 1 AMP at 110 

VAC. Use at least an 18 AWG wire.

NOTE: Configuration may be different

pending the analyzer ordered.

For AC-Powered Units

•  Use three wires. 
•  Connect them to the BLUE connector 

labeled L1, L2, and GND. 

For DC-Powered Units

•  Use two wires. 
•  Connect them to the BLUE connector  

labeled 12VDC or 24VDC.

•  Red wire to "+," and black wire to "-").

Содержание H2S/H2O/O2

Страница 1: ...grams PART 4 OPERATION 4 1 Screen Display 4 2 Calibration and Start up of the Analyzer 4 3 Calibrate Menu 4 3 1 H2S Calibration Menu 4 3 2 O2 Calibration 4 4 H2S Sensor Settings Menu 4 5 H2S Bump Test...

Страница 2: ...SOR WEAK Calibration has calculated sensor is approaching fail condition and will soon need replacement 2 Installation 2 1 Safety Requirements Operators are responsible for performing their own testin...

Страница 3: ...equire an additional pressure regulator at the sample tap One to three pressure drops may be necessary to ensure that the analyzer receives a sample that is free of liquids A typical system is illustr...

Страница 4: ...il signal A 5 AMP fuse is included to protect the analyzer from an overload A spare fuse is also provided The analyzer draws approximately 1 amp depending on its configuration If the analyzer is power...

Страница 5: ...section Sample Vent The sample sweeps through the instrument Sample gas flows here at all times Pipe the sample return to dispose of it wherever appropriate depending on the sample Remember To keep it...

Страница 6: ...100 degrees F The display will show when the heater is on The H2O readings may require approximately 24 48 hours to read accurately This time delay varies depending on the length of the sample line h...

Страница 7: ...nnet Drive Stafford TX 77477 281 240 4160 www tac controls com 3 Start up Operation Overview 3 2 Typical Flow Diagrams Typical H2S Flow Diagram NOTE Cal Valve not required when analyzer has a Manual C...

Страница 8: ...and pump and flow fail press ACKNOWLEDGED Green OK Yellow Sensor Weak Red Fail Red H2S Saturation Setpoints change to red and flashes during alarms 4 1 Screen Display After the analyzer is powered up...

Страница 9: ...pot on the front panel can be turned clockwise to increase the value and counter clockwise to decrease the value If necessary use a H2O detection tube to verify the amount of H2O at the sample point...

Страница 10: ...To Calibrate H2S 1 From the menu screen press the CALIBRATE button to enter the Calibrate menu screen 2 From the CALIBRATE menu screen press the CALIBRATE H2S BUTTON 3 Verify or change the H2S calibr...

Страница 11: ...N position 10 To return to the main screen press the READINGS button 4 3 2 O2 Calibration To Calibrate O2 1 From the CALIBRATE menu screen press the CALIBRATE O2 button 2 To verify or change the O2 ca...

Страница 12: ...hly equal The O2 reading and calibration gas should each read 25 ppm Formula Max Cts Full scale range x Cal Std Counts 32 000 100 x 25 8 000 32 000 100 x 50 16 000 Press the VERIFY CALIBRATION button...

Страница 13: ...2S SENSOR SETTINGS button 2 Press the respective sensor to turn it off 3 The main screen will show in red that the sensor is off To Perform a Bump Test 1 From the menu screen press the H2S BUMP TEST b...

Страница 14: ...Menu To Change the Alarm Settings 1 From the menu screen press the ALARM SETTINGS button 2 The alarm settings menu will be displayed Select the button for the setting you wish to change Set change th...

Страница 15: ...t the H2S levels remain higher than the range The analyzer may require purging Saturate mode can be changed to non latch However continuous saturations occurring in non latch mode may decrease sensor...

Страница 16: ...w Fail Assembly 12v or 24v To Clear Flow Fail 1 Ensure that the regulators and flow meter s are working properly 2 Check that the sample line for obstructions and clear them 3 Press the ACKNOWLEDGED b...

Страница 17: ...set change the 4 20 mA 3 Use the keypad to enter the values and press ENTER to save 4 Press the MENU button to return to the menu screen Press the READINGS button to return to the main screen 1 2 1 2...

Страница 18: ...ation 4 8 RTU Modbus Menu To Change the RTU Modbus Settings 1 From the menu screen press the RTU MODBUS button 2 Turn MODBUS OFF when configuring settings Press the desired button s to set change the...

Страница 19: ...the MENU button to return to the menu screen Press the READINGS button to return to the main screen To Change the Data Logging Settings 1 From the menu screen press the DATA LOGGING button 2 Press th...

Страница 20: ...the MENU button to return to the menu screen Press the READINGS button to return to the main screen To Access the Help Menu 1 From the menu screen press the HELP button 2 The help screen provides the...

Страница 21: ...WORD ENABLED or DISABLED to set The Password is set to 2030 To change the default password consult the TAC Support team 4 12 Password Protection Feature A new feature has been added to our analyzers P...

Страница 22: ...librate Sample Gas Pressure Read the sample gas pressure at the sample pressure gauge As mentioned in the installation section set this pressure to maintain the correct flow rate as determined during...

Страница 23: ...ation Step 1 Disconnect flow tubing at orange connecters x2 Step 2 D Step 4 Carefully lift Sensor Housing cover to expose H20 Sensor Set Sensor Housing safely to the side Step 1 Disconnect flow tubing...

Страница 24: ...lacing Membrane Filter To keep a clean sample to the analyzer it is important to replace the membrane filter as needed To Replace the Membrane Filter Unscrew the cap and pull the ring from the membran...

Страница 25: ...TEXAS ANALYTICAL CONTROLS 4418 Bluebonnet Drive Stafford TX 77477 281 240 4160 www tac controls com 5 Maintenance 11 15 1 4 11 3 8 7 5 8 9 15 16 1 4 11 5 8 15 1 4 7 5 8 16 3 8 9 14 1 2 5 6 Analyzer E...

Страница 26: ...6 1 Frozen Screen Run Idle Mode If the PLC freezes or gets stuck on the TAC Logo screen it maybe in the Idle mode To Change to Run Mode 1 Press the right corner of the touch screen to bring up the Sy...

Страница 27: ...Kbytes 1930752 Total Kbytes 1931264 6 2 PLC Loses Program Reload Program If the PLC loses the program it can be reloaded To Reload Program 1 Press the right corner of the touch screen to bring up the...

Страница 28: ...ghest quality To continue to improve our systems and add features and capabilities To protect the workers the environment and costly equipment We highly value our relationships with each customer Plea...

Отзывы: