background image

9

 

 

Batteries:

 

Batteries (4 x AA Alkaline) are supplied with the dispenser when new.  

 

Batteries will last in normal use for up to 12 months. 

 

 

Battery life can be checked using the following procedure:

 

 
 

 

1. 

Insert batteries in the dispenser.

 

 

 

2. 

Immediately watch the LCD display for the indication BATLEV.

 

 

 

3. 

At full charge a reading of between 1550 

-

 1600 (or more) will be shown.

 

 

4. 

At a reading of 1325 the battery warning indicator ‘’LOWBATT’’ will initiate.  

 

 

 5. 

At a reading of 1275 the unit will not operate, will blink ERROR and sound alarm 

sound 

04

 for 2 seconds, silence for 28 seconds, repeated. This is different to the  

standard alarm which sounds for 2 seconds repeating every 2 seconds.

 

 

NB:  1. 

Rechargeable batteries should not be used as they have an insufficient charge to    

reliably operate the unit.

 

 

         2. 

When the battery charge is reduced to a reading of 1325 the display will indicate   

LOWBAT. In normal operation the LOWBAT warning will be active for 1

-

2 months  

depending on how many daily doses are set. This gives early warning that a battery 
change is required.

 

Remember when changing batteries to insert them the correct way around (see figure 3. – p.3).

 

 

Removing Batteries:   

If the batteries are removed and replaced immediately the alarm 

settings are retained; only the clock would need to be reset if the batteries are not replaced 
immediately. 

 

 

Cleaning: 

 

 

Dispenser   

-

  

clean external surfaces of the dispenser body and lid with a disinfectant 

spray and / or wipe with a damp cloth. 

 

 

Internal tray  

-

  

the tray may be washed in warm water or in a dishwasher  (max. 35C).

 

 

Transport:

  If for any reason the dispenser is despatched by post, courier or other  carrier, it is 

important that 

all batteries

 are removed from the dispenser.

 

 

Important 

-

 Repositioning the Tray.  To avoid damaging the mechanism, the medicine tray 

must always be repositioned by lifting it out and then lowering directly into the case in the 
correct  position. Do not force the tray to rotate.

 

 

TabTimer cannot accept responsibility for faults or any other consequences arising from 
misuse  e.g. dropping, water damage, insertion of foreign objects, forced turning of the pill 
tray or forced opening  of the dispenser.  

 

Содержание Careousel MK3/11

Страница 1: ...TabTimer Careousel MK3 11 Automatic Pill Dispenser Instructions Note When opening for the first time the lid is unlocked The key and battery case securing screw can be found inside the dispenser...

Страница 2: ...TIC PILL DISPENSER The pack also includes Pharmacy label carrier and cover Three reversible Day Time discs Key Batteries 4 x AA Alkaline Battery case screw Lock Dispenser Tray Lid Display Buttons 1 2...

Страница 3: ...and pouring the pills into the hand or suitable container This will also cancel the alarm signal and the flashing light Technical data Dimensions length 190mm width 180mm height 56mm Weight 380g excl...

Страница 4: ...his section place the correct dosage in each section of the tray Note if a pill is inadvertently dropped in an incorrect compartment a pair of tweezers can be helpful to easily remove the pill Figure...

Страница 5: ...r the 2nd 2 alarm time and repeat the procedure for all subsequent alarm times required Press and hold down button 1 for a few seconds to exit the menu and revert to normal operation DISPENSER SETTING...

Страница 6: ...et at 00 00 We do not recommend that the BEFORE function is used for those with confusion AFTER DURATION The length of time the alarm will sound light will flash can be adjusted by the AFTER function...

Страница 7: ...ur HH then press 1 The time HH should be in advance of the first dose to be dispensed for the user If this function is not required do not set Note If at any time you go back to check the START settin...

Страница 8: ...g Procedure for pharmacy filling of the TabTimer Careousel MK3 11 please ask your pharmacist to contact TabTimer Pty Ltd Fault handling The alarm will sound continuously in the event of the dispenser...

Страница 9: ...aily doses are set This gives early warning that a battery change is required Remember when changing batteries to insert them the correct way around see figure 3 p 3 Removing Batteries If the batterie...

Страница 10: ...b screw in place to prevent access to the battery case Try to keep the dispenser reasonably horizontal If it is left upside down it will not rotate at the next alarm time and will give out the ERROR s...

Страница 11: ...all I can t get my fingers in to get the pills out also the alarm does not stop ringing The pills should be tipped out into the hand which also stops the alarm ringing There is an intermittent alarm s...

Страница 12: ...device and access more medicine than intended If there is any doubt that a patient will have an uncontrollable urge to access more medicine than prescribed then please do not use this device TabTimer...

Отзывы: