Pharmacy filling:

 

Many pharmacies will fill TabTimer® dispensers.  

 

The device meets the requirements of the Quality Care Pharmacy Program (QCPP)*.

 

 

For the suggested Standard Operating Procedure for pharmacy filling of the 

 

MK3/11 dispenser, ask your pharmacist to contact 
TabTimer® .

 

 

The new Pharmacy Filler tray 

is now available for 

easier Pharmacy filling of dispenser trays. 

 

 
 

Ask your Pharmacist to contact TabTimer® on 1300 822 846.

 

 

COMMONLY REPORTED ISSUES

 

 

 

‘I have just received my new dispenser but it is locked and there is no key.’

 

The unit is not locked but there is a knack to open (see Figure 2.) the key is inside.

 

 

‘The alarm sound/light has stopped working.’

 

Check the sound / light settings – they may have been inadvertently set to ‘00’

 

 

‘The dispenser alarmed during the night when it should not have’.

 

Check the settings, usually it has been incorrectly programmed to p.m. instead of a.m.  

 

Set to 24 hour clock to avoid this happening.

 

 

‘The dispenser didn’t allow for the clocks going back/forward.’

 

The date needs to be programmed into the unit. Check DLS is set to ‘01’.

 

 

‘The unit is not rotating.’

 

Check that there is no obstruction; check the battery level, check it is programmed correctly. 

 

Use the TEST function in Menu B.

 

 

‘The compartments are too small I can’t get my fingers in to get the pills out, also the 
alarm does not stop ringing.’

 

The pills should be tipped out into the hand, which also stops the alarm ringing.

 

 

‘There is an intermittent alarm sound and ‘ERROR’ is indicated on the screen.’

 

If the dispenser has been unable to rotate due to an obstruction it will show ‘ERROR’ on the display 
and emit the error alarm. This is a intermittent sound with a longer interval ( 30 seconds) compared to 
the normal ‘dispensing’ alarm interval. To stop alarm, take a battery out and then replace it.  

 

Clear the obstruction.  If the ERROR message continues there may be another fault in which case 
contact TabTimer® for inspection. 

 

 

‘The tray is rotating slowly.’

 

Check the battery level and replace if necessary. (Note

-

 Do not use rechargeable batteries, they do not 

have sufficient power for extended use).

 

 

‘I have dropped the dispenser, should I send it back for checking?’

 

Use the TEST function in menu B.  If all is operating satisfactorily then it is probably operating 

 

normally.  For extra reassurance reset the alarms and actually watch/listen to the rotation. 

 

 

‘If I take the batteries out will I have to reset all the alarms again?’

 

No, the alarm settings are retained, only the clock would need to be reset if the batteries are not 

 

replaced immediately. 

 

 

‘The dispenser is not working and ‘TILTED’ is displayed?’

 

Dispenser has been left upside down at the dosage time and all of the AFTER time.  

 

See FAULT HANDLING p.8. 

 

Содержание Careousel MK3/11

Страница 1: ...TabTimer Careousel MK3 11 Automatic Pill Dispenser Instructions Note When opening for the first time the lid is unlocked The key and battery case securing screw can be found inside the dispenser...

Страница 2: ...TIC PILL DISPENSER The pack also includes Pharmacy label carrier and cover Three reversible Day Time discs Key Batteries 4 x AA Alkaline Battery case screw Lock Dispenser Tray Lid Display Buttons 1 2...

Страница 3: ...and pouring the pills into the hand or suitable container This will also cancel the alarm signal and the flashing light Technical data Dimensions length 190mm width 180mm height 56mm Weight 380g excl...

Страница 4: ...his section place the correct dosage in each section of the tray Note if a pill is inadvertently dropped in an incorrect compartment a pair of tweezers can be helpful to easily remove the pill Figure...

Страница 5: ...r the 2nd 2 alarm time and repeat the procedure for all subsequent alarm times required Press and hold down button 1 for a few seconds to exit the menu and revert to normal operation DISPENSER SETTING...

Страница 6: ...et at 00 00 We do not recommend that the BEFORE function is used for those with confusion AFTER DURATION The length of time the alarm will sound light will flash can be adjusted by the AFTER function...

Страница 7: ...ur HH then press 1 The time HH should be in advance of the first dose to be dispensed for the user If this function is not required do not set Note If at any time you go back to check the START settin...

Страница 8: ...g Procedure for pharmacy filling of the TabTimer Careousel MK3 11 please ask your pharmacist to contact TabTimer Pty Ltd Fault handling The alarm will sound continuously in the event of the dispenser...

Страница 9: ...aily doses are set This gives early warning that a battery change is required Remember when changing batteries to insert them the correct way around see figure 3 p 3 Removing Batteries If the batterie...

Страница 10: ...b screw in place to prevent access to the battery case Try to keep the dispenser reasonably horizontal If it is left upside down it will not rotate at the next alarm time and will give out the ERROR s...

Страница 11: ...all I can t get my fingers in to get the pills out also the alarm does not stop ringing The pills should be tipped out into the hand which also stops the alarm ringing There is an intermittent alarm s...

Страница 12: ...device and access more medicine than intended If there is any doubt that a patient will have an uncontrollable urge to access more medicine than prescribed then please do not use this device TabTimer...

Отзывы: