Follow the steps below to attach the Upper Heated Plate to the AR 2000 rheometer head.
1.
Ensure that air at the correct pressure is supplied to the air-bearing, and remove the bearing cap. Turn on
the rheometer and raise the head to the maximum (use the
Head UP
button located on the instrument
control panel).
2.
Attach the Upper Heated Plate fixture to the mounting ring on the underside of the instrument head, using
the three captive screws provided. Note that the power cable should project to the right of the instrument
when viewed from the front, with the ports for the coolant and inert gas to the left (see Figure 7.2).
Figure 7.2
The Upper Heated Plate Shown Mounted on an AR 2000 Rheometer
3.
Disconnect the Peltier plate cable
from the Smart Swap™ socket,
using the “Release” button on the
instrument control panel.
4.
Connect the Peltier plate and
Upper Heated Plate cables to the
left and right sockets on the Smart
Swap™ Upper Heated Plate
adaptor respectively (see Figure
7.3 to the right).
Power Cable
Mounting Ring
Peltier Plate
Heating Element Cover
Upper Geometry
Coolant Ports
Figure 7.3
The Smart Swap™ UHP Adaptor
Connector to
Smart Swap™
Socket
Peltier Connector
UHP Connector
Содержание AR 2000
Страница 24: ...AR 2000 Operator s Manual 24...
Страница 30: ...AR 2000 Operator s Manual 30...
Страница 34: ...AR 2000 Operator s Manual 34...
Страница 66: ...AR 2000 Operator s Manual 66...
Страница 84: ...AR 2000 Operator s Manual 84...
Страница 120: ...AR 2000 Operator s Manual 120...
Страница 126: ...AR 2000 Operator s Manual 126...
Страница 128: ...AR 2000 Operator s Manual 128...
Страница 132: ...AR 2000 Operator s Manual 132...
Страница 134: ...AR 2000 Operator s Manual 134...