AR 2000 Operator's Manual
129
The LCD display on the front of the rheometer electronics box displays useful information and error messages.
Power On Messages
Immediately after power up of the rheometer, the display will show
'Initialising...'
After a few seconds this
will change to
'AR2000 7.xx xx/xx/xx' (x is version dependant)
. If it does not, there is a problem with the
rheometer.
After a few more seconds, the display will either show
'System test Ok'
or
'System test failed Y,'
where Y is an
error code as show below:
1
ROM checksum error
Either bad firmware or
Try reloading firmware, otherwise call for
or hardware fault
service.
2
RAM error
Hardware fault
Call for service.
4
Dual port RAM error
Hardware fault
Call for service.
40
Battery failure
Low battery
Sometimes seen the first time the system is
restarted after a firmware upgrade.
Cycles power off then on.
80
Backup RAM error
Hardware fault
Call for service.
400
Parameter block
Either corrupted
Call for service.
checksum error
internal system
parameters or
hardware fault
Initialising ...
During start up of the rheometer, the following items are shown as initializing:
•
Electronics
•
Power board
•
Instrument
Bearing overspeed
Shown when the rheometer bearing rotation exceeds the specified maximum speed.
Appendix D
LCD Display Messages
Содержание AR 2000
Страница 24: ...AR 2000 Operator s Manual 24...
Страница 30: ...AR 2000 Operator s Manual 30...
Страница 34: ...AR 2000 Operator s Manual 34...
Страница 66: ...AR 2000 Operator s Manual 66...
Страница 84: ...AR 2000 Operator s Manual 84...
Страница 120: ...AR 2000 Operator s Manual 120...
Страница 126: ...AR 2000 Operator s Manual 126...
Страница 128: ...AR 2000 Operator s Manual 128...
Страница 132: ...AR 2000 Operator s Manual 132...
Страница 134: ...AR 2000 Operator s Manual 134...