Operation Manual

TW-2B

gauge is calibrated. Should the aberration be larger than 0 +/-10µm, please put the sensor once 
more against the green side of the calibration sheet, wait until the result stabilizes and press the

 

OK [MENU] 

button while the sensor still touches the sheet. On the display will appear a 

OK...

 

caption and the device will return automatically to the main menu.

If  the  device  is  calibrated,  You  can  cancel  the 

Zero

  function  by  pressing  the

  OK  [MENU] 

button. Please note not to press the

 OK [MENU] 

button before the 

------

 caption appears in 

the second line of the display – otherwise the gauge will decalibrate!

WARNING! 

W

hIle

 

zeRoING

The

 

cAlIbRATIoN

 

sheeT

 

should

 

be

 

plAced

 

oN

 

A

 

flAT

sTAble

 

NoN

-

meTAlIc

 

suRfAce

 (

e

.

G

do

 

NoT

 

plAce

 

The

 

sheeT

 

oN

 

The

 

bodyWoRk

). A

lso

pleAse

 

AvoId

 

holdING

 

The

 

sheeT

 

IN

 

y

ouR

 

hANds

 

As

 

ThIs

 

mAy

 

AffecT

 

The

 

opeRATIoN

.

Please note that zeroing is worth to be made when the measurement conditions are changing 
(e.g. the temperature or humidity rises or falls).

WARNING! 

1. u

se

 

oNly

 

AlkAlINe

 

bATTeRIes

 

As

 

NoRmAl

 

bATTeRIes

 

pRovIde

 

uNsuffIcIeNT

 

poWeR

. 2. I

NsTeAd

 

of

 

A

 

AlkAlINe

 

bATTeRy

 y

ou

 

cAN

 

use

 

A

 

compATIble

fully

 

chARGed

 

Accu

. 3. I

f

 

The

 

devIce

 

seems

 

To

 

WoRk

 

ImpRopeRly

 

IT

 

cAN

 

be

 

cAused

 

by

 

AN

 

empTy

 

bATTeRy

 

oR

 

Accu

.

To take a measurement please put the device’s sensor against the examined surface. The sensor 
should adhere as flat as possible. The examined surface should be clean and smooth, otherwise 
the measurement may be incorrect. It is advised to hold the gauge with both hands and press 
gently  towards  the  examined  surface  to  eliminate  vibration.  Usually  it  takes  1-2  seconds  to 
stabilize the measurement result.

Gauge manual

3. 

The device is operated with 2 buttons:

button 

• 

OK  [MENU]

:  turns  the  device  on,  accepts  the  selected  functions  and  returns 

from them to the main menu, when pressed during the measurement it memorizes the 
displayed value;
button 

• 

FUNCTION

: switches between the functions in the main menu, switches between 

the stored measurement values.

To turn the device on please press the

 OK[MENU] 

button for about 1 second. After displaying 

the company’s logo and the device’s type, the device switches automatically to the main menu. 
You will see on the display 

Measur

 in the first line and 

Memory

 in the second line. The 

*

 

symbol indicates the currently selected function. 

Содержание TW-2B

Страница 1: ...Thoma Tel Fax 48 95 741 11 62 info t4w eu www t4w eu Coating Thickness Gauge TW 2B Operation Manual art nr 5924 7 ...

Страница 2: ...peration Manual TW 2B Table of contents 1 Technical specifications 3 2 Getting started 3 3 Gauge manual 4 Guarantee TW 2B 8 The price of this product includes the recycling fee of 0 10 PLN Before You start to use the device please read this operation manual ...

Страница 3: ...age much easier Before You start to use the device please place a new alkaline battery or a fully charged accu in the back side of the device s housing In order to do it please take off the battery flap and place correctly the battery or accu by connecting it to the clips on the cable Before You start to take measurements please place a new alkaline battery or a fully charged accu in the back side...

Страница 4: ...normal batteries provide unsufficient power 2 Instead of a alkaline battery You can use a compatible fully charged accu 3 If the device seems to work improperly it can be caused by an empty battery or accu To take a measurement please put the device s sensor against the examined surface The sensor should adhere as flat as possible The examined surface should be clean and smooth otherwise the measu...

Страница 5: ... select a function please press the OK MENU button Measure function Measur After You select this function the display will show Measur in the first line and um in the second The gauge is ready to take measurement After You put the sensor against the body of the car the result of the measurement will be shown in the second line To store the result please press the OK MENU button when taking the mea...

Страница 6: ...eel Fe You can switch between the sheet materials with the FUNKTION button to accept the selection press the OK MENU button Delete function Delete Choose this function to delete all measurement values stored in the device s EEPROM internal memory After deleting the stored results the device returns to the main menu automatically Zeroing function Zero Use this function before You start to take meas...

Страница 7: ...pears in the second line of the display otherwise the gauge will decalibrate WARNING While zeroing the calibration sheet should be placed on a flat stable non metalic surface e g do not place the sheet on the bodywork Also please avoid holding the sheet in Your hands as this may affect the operation Please note that zeroing is worth to be made when the measurement conditions are changing e g the t...

Страница 8: ... be shipped to the Producer it happens on the shipper s responsibility and cost 8 The device will be not accepted by the Service if pt 6 is not kept if the defect is of no workmanship or 9 material nature if the warranty card is not filled out or the device is delivered with broken seal This guarantee shall not apply to damage caused through fire accident misuse incorrect adjustment or 10 repair i...

Отзывы: