background image

  ROW weighing scales operation manual                                           

REV: C1, June 2015

                                              ROW weighing scales operation manual 

 

   

accumulation function is not automatic.   

Secondary function ( 

C  or     

) , is to act as a clear key   or to 

move active digits left when setting values for parameters or other 

functions. 

 

UNIT or U 

Press this key to select the weight unit.   

Secondary function (ESC) , is to  return to normal operation when 

the scale is in a parameter setting mode. 

 

ON/ OFF   

Turn on or off the power. 

 

3. BASIC OPERATION 

3.1 Zeroing The Display 

You can press the 

ZERO

 key at any time to set the zero point from 

which all other weighing and counting is measured, within 4% of power 

up zero. This will usually only be necessary when the platform is empty. 

When the zero point is obtained the display will show the indicator for 

zero. 

The scale has an automatic rezeroing function to account for minor 

drifting or accumulation of material on the platform. However you may 

need to press the 

ZERO

  key to rezero the scale if small amounts of 

weight are shown when the platform is empty. 

 

3.2 Taring 

Zero the scale by pressing the 

ZERO

  key if necessary. The zero 

indicator will be on. 

Place a container on the platform, a value for its weight will be 

displayed. 

 

-2- 

 

by using the keys 

PRINT/M+

G/N

TARE

 to select the required setting. 

Enter the selected setting by pressing 

ZERO

, display will be 

shown ”

load

”. 

Load the calibration mass weight on the platform and wait few 

seconds for display stability. 

After the stable indicator on press 

ZERO

, display will be 

shown ”

pass

”.   

After the calibration the display will start a self test. Remove the load 

from platform during the  test. Display will come to weighing mode 

automatically. 

If display will be shown any error or incorrect value, repeat the 

procedure again. 

 

8. BATTERY OPERATION          

The weighing scale can be operated from the battery if desired. The 

battery life is approximately 35 hours. 

When the battery needs charging, the battery symbol will blank. The 

scale will still operate for about 1 hours after which it will automatically 

switch off to protect the battery. 

To charge the battery simply plug into the AC adapter. The scale 

does not need to be turned on. 

The battery should be charged for 12 hours for full capacity. 

In the LCD display has an battery indicator to indicate the status of 

battery charging.  When the scale is plugged into AC adapter,  the 

internal battery will be charged.  When turn on the indicator, if the 

battery  indicator  is  full then  the battery has a full charge.  If it is half 

then  the battery is nearly discharged and empty  indicates the battery 

should be charged. When charge the battery, the indicator will be flicker. 

As the battery is used it may fail to hold a full charge. If the battery 

life becomes unacceptable then contact your distributor. 

 

 

 

-7-

 