![SYSTRONIK MULTILYZER STe Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/systronik/multilyzer-ste/multilyzer-ste_instruction-manual_1423597048.webp)
48
MULTILYZER
®
ST/STe
Connect the gas system with the MULTILYZER
®
STe.
The gas system must now be connected to the MULTILYZER
®
STe.
Turn on and allow the appropriate pressure into the system.
With "START Pressure Loss." The pressure drop measurement is
started. A counter in the main display shows the current running time
in seconds and minutes.
After the set elapsed time duration a signal sounds and in the infor-
mation line the message "STOP pressure loss" .The elapsed meas-
urement time is also displayed. The readings are held in the main
display, and you can then print the test report or save it to the Mi-
croSD Card.