 Centre for Biological Engineering 

 

Standard Operating Procedure  

 

 

 

 

 

 

 

SOP025 

 

Title:  USE AND MAINTENANCE OF THE SYSTEC VX 95 AUTOCLAVES

 

 

Location: 

Autoclave Room, H31, CBE 

 

 

Version 004 

Effective Date: 28

th

 June 2020 

Review 28

th

 June 2022 

 
Written by: A. Chandra  
 

Reviewed by: P.Hourd 

Approved by: R.I.Temple 

 
 

Page 19 of 26 

 

Mastercopy

 

 

 

 

 

CAUTION:

 There is danger of crushing and or amputation between the door and the edge of the 

autoclave chamber if the door has not been completely opened.  

  

(iii) 

If required, remove the flexible temperature sensor from the autoclaved item (or 

reference vessel) and fix it to the holder provided for it in the door.  

  

 

(iv)  

If fully loaded tubs/baskets are too heavy to lift manually they should be 

partially unloaded 

 

before removal from the autoclave. Alternatively an electrically-operated lifting device is available 

(Max load = 35Kg) to assist unloading. It is mounted directly onto the appliance. It enables you to 
load and unload heavy loads without any exertion. A swivel arm also facilitates operation.  

  

 

(v)  

Remove the basket, tub or item from the autoclave chamber and check the indicator tape:   

  

  Totally black lines indicates that steam exposure was effective  

  Not totally black lines indicate failure during the discard cycle. Re-autoclave the waste 

by loading with sterilisation indicators and record the results in the User Log  

 If it is fails 

again, report the problem to the responsible person or laboratory manager  

 

 

5.9. Switching off the autoclave  

  

(i) 

After use, switch off the autoclave at the main switch located at the side of the autoclave  

  

(ii) 

If some danger has been detected in the autoclave switch off the autoclave at the main 

switch.   

  

NOTE: 

When the power supply is switched off during operation, the current process is 

interrupted. The autoclave gradually returns to a depressurized state and low temperature, and 
can then be opened safely. After the power supply is interrupted and then switched on again, 
the autoclave returns to the state it was in before the power supply was interrupted. Any error 
messages that may have been displayed on the control panel before the interruption are 
displayed exactly as before. The pressure and temperature values displayed reflect the current 
state.   

  
(iii) 

In case of emergency; to shut off the steam source switch the power off using the 

emergency safety switch located on the ceiling of the autoclave room above the autoclaves 
(

NOTE: 

a pole is provided to enable the operator to perform this action). Leave the autoclave 

room and prevent others from entering. Consult the Laboratory Manager or Responsible Person 
and contact the DSO.  

  

Содержание CBE045

Страница 1: ...t not waste that may include bottles beakers pipettes forceps tubing etc All procedures described in this SOP for solid or liquid waste treatment have been validated Typical waste batches are describe...

Страница 2: ...it must be ensured that the appliance is properly switched off by means of the main switch on the side of the autoclave vii Risks due to wear The autoclave must be serviced at the regular intervals a...

Страница 3: ...arameters described in the manual and this may result in danger for the operating personnel or impair the sterilization result Changes MUST only be made by trained personnel and MUST be documented in...

Страница 4: ...n MUST ensure that the University Estates and Buildings Department is notified of all autoclave installation They will inspect the autoclave at the statutorily required interval Notification of such i...

Страница 5: ...AND MATERIALS 1 Systec VX 95 Autoclaves Number CBE045 and CBE 044 2 Printer 3 De ionised water vessel 4 Autoclave Basket 5 Autoclave Bucket item J807 from nisbets co uk 6 Integrated hoist 7 Autoclave...

Страница 6: ...hes the vertical stop DO NOT pull the door from the front v Ensure that the flexible temperature sensor PT100 is placed in the holder in the door when the door is opened HINT It is a good idea to put...

Страница 7: ...N A YES Solid Laboratory waste melting items in Bags in steel bucket in basket 121 C 15 min 99 C 5 N A YES Liquid Laboratory waste in bottles max 5L with probe in 2L reference bottle 121 C 15 min 80...

Страница 8: ...cycle MUST be used for sterilisation of liquid waste Liquid waste must be kept in an autoclavable container The container MUST NOT be tightly sealed A Reference volume blank containing 2L of distille...

Страница 9: ...unloading of the autoclave an electrically operated lifting hoist is available Max load 35Kg It is mounted directly onto the appliance It enables you to load and unload heavy loads without any exertio...

Страница 10: ...lly operated lifting device is available Max load 35Kg It is mounted directly onto the appliance It enables you to load and unload heavy loads without any exertion A swivel arm also facilitates operat...

Страница 11: ...litates operation v DO NOT place the flexible temperature sensor in the item being autoclaved ie the item may melt The temperature sensor should be placed in the holder provided for it in the steriliz...

Страница 12: ...ads without any exertion A swivel arm also facilitates operation vi Place autoclave indicator tape white banded tape on outside of the bottle vii DO NOT leave bottles for autoclaving in the autoclave...

Страница 13: ...ifting device is available Max load 35Kg It is mounted directly onto the appliance It enables you to load and unload heavy loads without any exertion A swivel arm also facilitates operation vi Label t...

Страница 14: ...is recommended that this cycle is validated i CYCLE 11 Liquids with an operating cycle fixed at 121 C and 15 minutes MUST be used for the sterilisation of liquids in suitable OPEN vessels without act...

Страница 15: ...erilized are OPEN under no circumstances vessels should be tightly shut Cap medium solution bottles loosely whether empty or filled to prevent explosions due to expansion If using Cycles 8 9 or 10 if...

Страница 16: ...ds cycles causes danger through boiling delays or exploding vessels and can result in serious injuries To prevent liquids mistakenly being autoclaved in a solids cycle a safety query is performed afte...

Страница 17: ...up or sterilization phase causes the cycle to pass directly to the cool down phase In this case to guarantee complete sterilization you must repeat the entire sterilization process By interrupting the...

Страница 18: ...d to scalding iii To open the door press the Open button displayed on the front panel The autoclave will alarm for 30 seconds to warn the operator to move back The operator should stand clear The door...

Страница 19: ...ines indicate failure during the discard cycle Re autoclave the waste by loading with sterilisation indicators and record the results in the User Log If it is fails again report the problem to the res...

Страница 20: ...og sheet provided The printed record that documents parameters while a cycle is running and provides cycle data at the end of the cycle MUST be detached and the Log number from the template form writt...

Страница 21: ...sight of or copies of records during visits to the University These should be retained in the Equipment File 5 12 Preventive Maintenance Measures The maintenance and care tasks described in this secti...

Страница 22: ...ager or Responsible Person Once a month the Responsible Person MUST carry out the cleaning cycle Cycle 12 for cleaning the autoclave To do this the autoclave heats up to a temperature of 134 C and ste...

Страница 23: ...o a temperature of 134 C and sterilizes the interior for one minute Record the cycle run in the autoclave log v Uncontaminated cracked glassware must be disposed of properly in the Broken Glass dispos...

Страница 24: ...be carried out under a Permit to Work scheme v When a fault occurs during a make safe cycle an assessment of risk should be made and appropriate action taken It may be necessary to disinfect those ch...

Страница 25: ...5000ml pipette tips 1 x pipette tip box 3 x pipette tip box packaging 10 x large gloves Cardboard glove box 3 x T 175 flasks with bags 5 x 25 ml centrifuge tubes 5 x 20ml centrifuge tubes 10 x bijous...

Страница 26: ...9 which are a 5 Make Safe for solid waste which can melt and needs secondary containment in steel buckets b 6 Liquid Make Safe where maximum liquid is 5L 2 Section 7 added note regarding PPE to be wor...