Automated Cleaning Instructions.

For Synthes Trauma Recon System.

3

Remove debris

Rinse device under running cold tap water for a 
minimum of two minutes. Use a sponge, soft lint-free
cloth and/or soft-bristled brush to assist in the removal of
gross soil and debris. Clean all cannulations, handpiece
and attachments, with the Cleaning Brush (516.101).

Manipulate all moving parts such as the triggers, release
sleeves for attachments, mode switch, etc. under running
cold tap water to loosen and remove gross debris. 

Note:

Do not use pointed objects for cleaning.

Automated Cleaning Instructions with Manual Precleaning.

4

Spray with solution

Spray and wipe device using a neutral pH enzymatic 
solution or foam spray for a minimum of two minutes.
Follow the enzymatic detergent manufacturer’s 
instructions for use for correct temperature, water 
quality and concentration/dilution.

6

Rinse with tap water

Rinse device thoroughly using running cool to lukewarm
tap water for a minimum of two minutes. Use a syringe,
pipette or water jet to flush lumens and channels. Actuate
joints, handles and other moveable device features in 
order to rinse thoroughly under running water.

2

Return batteries to charger

Return batteries/power modules to charger after 
each use.

Important: Do not sterilize batteries or power
modules! Batteries/power modules will no 
longer function.

This information is provided in accordance with ISO 17664 and ANSI/AAMI ST81. 

1

Disassemble

Disassemble device if applicable, and remove the battery
or power module from casing or handpiece.

Batteries/power modules and charger may be wiped
with a cloth. DO NOT disinfect, immerse, or sterilize 
batteries/power modules.

Note: 

The screw coupling of the Sagittal Saw 

Attachment, long, 05.001.224, completely 
disassembles from saw for cleaning

.

Visually inspect device. Repeat Steps 3–6 until no visible soil remains. 
Note: Manual precleaning as described above must be followed by the 
mechanical cleaning procedure.

– Mechanical cleaning/disinfection is an additional 

stress for power equipment, especially for seals and
bearings. Therefore, systems must be properly 
lubricated and serviced regularly.

Note:

Refer to back of chart for automatic washing instructions.

Important:

– Manual precleaning prior to mechanical cleaning/

disinfection is important to ensure cannulations and
other difficult to access sites are clean.

– Alternative cleaning/disinfection procedures other than

the procedure described below (including manual 
precleaning) have not been validated by Synthes.

5

Clean with detergent

Clean device manually under running warm water using
a neutral pH enzymatic detergent for a minimum of five
minutes. Manipulate all moving parts under running 
water. Use a soft-bristled brush and/or soft lint-free 
cloth to remove all visible soil and debris. Follow the 
enzymatic detergent manufacturer’s instructions for 
use for correct temperature, water quality and 
concentration/dilution.

Note:

For Sagittal Saw Attachment, long, 05.001.224,

process the Screw Coupling in an ultrasonic cleaner 
using a neutral pH enzymatic cleaner for a minimum 
of 15 minutes.