Synthes
35
Freehand Distal Locking— Long Nails (300 mm to 460 mm)
1
Reconfirm reduction
Instrument
511.30* Radiolucent Drive
Confirm reduction of the fracture with AP and lateral images.
For static interlocking, use the superior locking hole. For
dynamic interlocking, use the inferior locking hole only.
2
Align image
Align the image intensifier with the hole in the nail until
a perfect circle is visible in the center of the screen.
Important:
Confirm that the nail is securely connected to
the insertion handle, especially after hammering.
* Also available
Round (correct)