32
Synrad
Firestar OEM v30
reference guide
Troubleshooting
(cont.)
Symptom:
■
The
PWR
indicator blinks three times, pauses 1/2 second and repeats. Both
RDY
and
LASE
LEDs are Off. On the
Interface A
(DB-9) connector,
Lase Indicator
,
Laser Ready
, and
Overtemp Fault
outputs are Low, but the
DC Voltage
Fault
output is High. On the
Interface B
(RJ45) connector, the
Temperature Ok
signal is High while the
Laser Ok
and
Voltage Ok
outputs are Low.
Possible Causes:
■
A DC Sense fault has occurred.
Remove DC power from the laser, wait 30 seconds, and then re-apply DC power. If the fault fails to clear, contact SYNRAD
or a SYNRAD Authorized Distributor.
Symptom:
■
Both
PWR
and
RDY
indicators are flashing. The
LASE
LED is Off. On the
Interface A
(DB-9) connector, all outputs
(
Lase Indicator
,
Laser Ready
,
DC Voltage Fault
, and
Overtemp Fault
) are Low. On the
Interface B
(RJ45) connector,
the
Temperature Ok
and
Voltage Ok
outputs are High while the
Laser Ok
signal is Low.
Possible Causes:
■
A PWM Sense fault has occurred.
Remove DC power from the laser, wait 30 seconds, and then re-apply DC power. If the fault fails to clear, contact SYNRAD
or a SYNRAD Authorized Distributor.
DB-9 connector
Symptom:
■
The built-in +5 VDC, 250 mA voltage output on Pin 5,
DC Out
, has been accidentally shorted to ground.
Possible Causes:
■
In the event that Pin 5,
DC Out
, is shorted to ground, the OEM v30 laser will not be damaged as the voltage regulator
is protected by a current-limiting circuit.
RJ45 connector
Symptom:
■
The built-in +15 VDC, 250 mA voltage output on Pin 2,
DC Out
, has been accidentally shorted to ground.
Possible Causes:
■
In the event that Pin 2,
DC Out
is shorted to ground, the OEM v30 laser will shutdown, but restarts automatically
once the short-circuit is removed.
Beam delivery optics
Symptom:
■
The laser appears to lose power over time; laser output power must be increased to maintain previous performance.
Possible Causes:
■
Beam delivery optics are coated by vapor residue or debris.