Red coloured shading indicates overexposure of white. Blue coloured shading indicates 

underexposure. The optimum exposure time is one showing no red shading.

The InGenius LHR has a very wide dynamic range which means that higher levels of light can be 

detected by the camera before it reaches saturation point.
To check for saturation on a grabbed image, click on the Histogram icon.

A graph will be displayed indicating which grey shades have been detected by the camera, from 0 

- 4095 (or 0-65535 if EDR has been used). Providing that all areas of the image are within this 

range and within the linear range of the camera, quantification may be performed with confidence.

Please note that for dyes that have a lower dynamic range compared to the camera  e.g. silver or 

coomassie then it will not be possible to utilise the full dynamic range of the camera.

16) Auto exposure

17) Printing

18) Saving an image

19) UV safety override

Auto exposure

- InGenius LHR has a unique auto-exposure system leaving the software to decide 

the optimum camera settings to "grab" the best image. The image will have no saturated areas and 

is fully quantifiable. This setting should only be used when capturing images of fluorescent 

samples. It is especially useful when the gel is dark or if the bands are very faint. To activate this 

function simply click on the AUTO EXPOSE button. 

Depending upon the gel this can take several seconds until the camera has found the optimum 

settings. (A progress indicator shows the automatic exposure control is operating).

Once it has completed the process the "grabbed" image is displayed in the image window and the 

lighting is automatically switched off.

Printing

- When an image is "grabbed" the PRINT icon is available.

Click this to send a copy of the image to the thermal printer 

for a hard copy output.

Saving the image

- You have the choice of either selecting FILE/SAVE AS or FILE/EXPORT. In 

the first case the image can be saved in the Syngene secure SGD format for true GLP. With the 

second option, the image can be saved in a range of formats including TIF, BMP, GIF, PCX, TGA, 

WPG, and JPEG. Images can then be exported to other software packages.

For more detailed information concerning capture and enhancement including custom user 

configurations, annotation, filter and colour overlays, image browser etc. - please refer to the full 

manual supplied.

UV Safety override 

-To use the transilluminator with the door open override the UV safety 

switch by pulling down the safety interlock switch which is located above the door opening on the 

right.

Note

- Please take adequate precautions to ensure that the UV light cannot reach your eyes or 

skin or anyone else’s by wearing the appropriate protective equipment. 

Note

- The interlock will be re-set when the door is closed.

* If both the 

focus indicator

and 

increased effective pixel

features are not present then install 

GeneTools software as these features are licensed with the analysis software.

S Y N G E N E

A  DIVISION  OF  THE  SYNOPTICS  GROUP

CONTACT SYNGENE:

EUROPE

IN USA:

BEACON HOUSE

5108 PEGASUS COURT, SUITE M

NUFFIELD ROAD

FREDERICK 

CAMBRIDGE

MD 21704

CB4 1TF

Tel: +44 (0)1223 727123

Tel: 800 686 4407/301 662 2863

Fax: +44 (0)1223 727101

Fax: 301 631 3977

Email: [email protected]

Email: [email protected]

Email: [email protected]

Email: [email protected]

www.syngene.com

4

IGLHRINST0408

Содержание InGenius LHR

Страница 1: ...port on the PC until the GeneSnap Camera drivers have been installed Connect electrical power leads to each component but DO NOT SWITCH ON AT THIS TIME Switching on with the Syngene supplied PC Attach the free end of the camera USB cable to a USB 2 port on the PC Ensure it does not share a paired port with a slower device for instance a mouse otherwise this will compromise the response time of the...

Страница 2: ...Image capture and enhancement is carried out within GeneSnap if the program does not open automatically simply double click on the GeneSnap desktop icon You will now see the GeneSnap User box Type into the space your name and click the OK button This is now your configuration and any changes you make to the way you use or set up the program will be stored as defaults under your name Each user can ...

Страница 3: ...ab the image The camera icon will turn GREEN To return to a live image simply click the GREEN camera icon and it will return to RED and the camera is again showing the current sample Further controls Exposure time Above we grabbed an image in real time using the standard exposure setting of the camera However in some instances you may wish to extend the exposure time to resolve fainter bands In th...

Страница 4: ...abbed image is displayed in the image window and the lighting is automatically switched off Printing When an image is grabbed the PRINT icon is available Click this to send a copy of the image to the thermal printer for a hard copy output Saving the image You have the choice of either selecting FILE SAVE AS or FILE EXPORT In the first case the image can be saved in the Syngene secure SGD format fo...

Отзывы: