To change the Parallel BCD test mode selection, send the following send the following string to the
Serial/Network port:
F78<SP>B<N><SP>TEST<SP><SEL>
where:
F
= ASCII character F.
78
= function number.
<SP> = ASCII space character.
B
= ASCII character to denote Option Bay number follows
<N>
= Option Bay Number, 1 through 10.
TEST = ASCII character to denote test mode selection.
<SEL> = ENABLE or DISABLE.
<LT> = line terminator; for output strings (a carriage return and line feed ) or for input strings (a
carriage return only).
Sample entry: F78 B2 TEST ENABLE<CR>
Response: OK<CR><LF>
The Serial/Network port will respond with the message “ERROR 01 VALUE OUT OF RANGE” if the
input string was in the correct format but contained a value, probably numeric, that was out of the range
of acceptable values.
The Serial/Network port will respond with the message “ERROR 02 SYNTAX” if it receives a string in an
incorrect format.
The Serial/Network port will respond with the message “ERROR 03 BAD/MISSING FIELD” if the input
string lacks a required field.
The Serial/Network port will respond with the message “ERROR 04 BAD DATA / TIMEOUT
CONDITION” if option card does not respond to the XLi fast enough. Re-enter the command to receive
the data.