Appendix A Performance Tests
5071A Complete Performance Tests
Operating and Programming Manual
147
4
On the 5071A, set output ports 1 and 2 to 5 MHz using the front panel controls.
5
Repeat step 3.
6
Connect 5071A Port 2 to the 3585B Spectrum Analyzer and repeat steps 2
through 5 to test Port 2.
7
Record the actual reading in the appropriate place in the Performance Test
Record.
B. Non-Harmonic (Spurious) Signal Check for the 5 and 10 MHz
Outputs
Non-harmonically related (spurious) signals on the 5 and 10 MHz outputs must
be more than 80 dBc below the output signal levels. To perform this test, a
spectrum analyzer is tuned to the 5 or 10 MHz signal and an amplitude reference
is established. The output frequency spectrum is then examined at 1 MHz on
either side of the center frequency to determine the fundamental-to-sideband
amplitude relationship for any signals occurring within this range.
Equipment
HP/Agilent 3585B Spectrum Analyzer
Setup
Figure A-4. Non-Harmonic Distortion Test Setup
Содержание 5071A
Страница 2: ......
Страница 28: ...Preface xxvi Operating and Programming Manual ...
Страница 29: ...____________________________________________ Chapter 1 Getting Started Getting Started ...
Страница 32: ...Chapter 1 Getting Started The 5071A at a Glance 4 Operating and Programming Manual ...
Страница 33: ...Chapter 1 Getting Started The 5071A at a Glance Operating and Programming Manual 5 ...
Страница 34: ...Chapter 1 Getting Started The 5071A at a Glance 6 Operating and Programming Manual 48V OPTION REAR PANEL VIEW ...
Страница 40: ...Chapter 1 Getting Started Performing Basic Tasks 12 Operating and Programming Manual ...
Страница 41: ...____________________________________________ Chapter 2 Using the 5071A Using the 5071A ...
Страница 57: ...____________________________________________ Chapter 3 In Case of Difficulty In Case of Difficulty ...
Страница 97: ...____________________________________________ Chapter 5 Command Reference Command Reference ...
Страница 119: ...Chapter 5 Command Reference Description of SCPI Commands Operating and Programming Manual 91 ...
Страница 136: ...Chapter 5 Command Reference Error Messages Codes 108 Operating and Programming Manual ...
Страница 137: ...____________________________________________ Chapter 6 Specifications Specifications ...
Страница 145: ...____________________________________________ Chapter 7 Installation Installation ...
Страница 162: ...Chapter 7 Installation Cesium Beam Tube Replacement 134 Operating and Programming Manual ...
Страница 163: ...____________________________________________ Appendix A Performance Tests Performance Tests Verifying Specifications ...
Страница 186: ...Appendix A Performance Tests 5071A PERFORMANCE TEST 158 Operating and Programming Manual ...
Страница 187: ...____________________________________________ Appendix B Glossary Glossary ...
Страница 190: ...Appendix B Glossary 5071A Glossary 162 Operating and Programming Manual ...