Chapter 6 Specifications
Supplemental Characteristics
114
Operating and Programming Manual
Environmental
Temperature:
Operating: 0 to 55 C.
Non-operating: -40 to 70 C.
Humidity:
0 to 95% relative humidity (at
maximum temperature of 45 C).
Atmospheric pressure:
<1 x 10
-13
change in frequency
for pressure change down to 19
kPa (equivalent to an altitude of
about 12.2 km).
Shock and vibration:
MIL-T-28800D, Type III,
Class 5.
Hammer Blow Shock Test,
MIL-S-901C, Grade A, Class 1,
Type A.
EMI: C1SPR 22, Class A.
CISPR 11, Group 1, Class A;
MIL-STD-461 C, Part 7, Class B.
dc magnetic field: up to
7.8 Gauss.
Weight:
Net weight: 30 kg
Shipping weight: 40 kg
Dimensions
Height
133.4 mm
Width
425.5 mm
Depth
523.9 mm
This includes spacers which extend
25.4 mm beyond the rear panel.
Warranty
Electronics
2 years.
Cesium beam tube
Standard tube: 10 years.
High-performance tube
(Option 001): 5 years (serial prefixes
US4240 and above)
Accessories Furnished
Power Cord
180 cm detachable.
DC input Connector:
Part number 1251-0126
(5 contact), Cannon MS3106E-14S-
5S (series ME).
Cabinet front handles kit.
Accessories Available
Rack Slides: (see Options for other
rack mounting hardware)
Fixed Slides: order part number
1494-0059.
Application note available
Application Note 1289,
The
Science of Timekeeping.
5071A Options
High Performance Cesium Beam
Tube.
300
: Blank Front Panel. This
option deletes the LED time
display, the user keypad
interface, the 40-character LCD
display, and the front panel door.
The Continuous Operation and
Attention indicators are still
present, and one each BNC
connector for 1 PPS output and
Sync input are still present. All
communication and control of
the 5071A is done via the rear-
panel RS-232 (serial) port.
046
: This option removes the
internal 45-minute standby
batteries. The 5071A will only
operate on ac mains or external
n24 Vdc power. This
option is not compatible with
48V option.
048:
This option equips the
frequency standard to operate
from 48-Vdc nominal power.
The range that the frequency
standard will accept is 40-58
Vdc. Maximum dc power
consumption is 100W. Two
independent 48-Vdc power
sources may be supplied. The
power sources must be of the
same polarity but either
polarity may be used. The
instrument automatically uses
48-Vdc source if one is present.
If no dc source is available, the
instrument will attempt to use
ac power. Unit will draw dc
power in preference to ac
power. The internal battery
normally supplied with the
5071A is not available when
48V option is installed.
Damage level: 72 V.
Maximum allowable voltage
between either input terminal and
chassis is 60V.
5071A Options (continued)
908:
Rack Flange Kit. Use for
mounting the 5071A when
front handles will not be used.
Also available under part
number 5062-3977.
WARNING: Do not mount
this product by the rack
flanges alone. Always use a
shelf, rails, or slides.
913:
Rack Flange Kit. Use with
front handles supplied with
5071A. Also available under
part number 5062-3983.
WARNING: Do not mount
this product by the rack
flanges alone. Always use a
shelf, rails, or slides.
W30:
Two additional years of
―Return to Symmetricom‖ service
on the electronics (the cesium
beam tube is warranted
separately). This option adds to
the basic electronics warranty to
provide a total of three years of
coverage.
W50:
Four additional years of
―Return to Symmetricom‖ service
on the electronics (the cesium
beam tube is warranted
separately). This option adds to
the basic electronics warranty to
provide a total of five years of
coverage.
† Footnotes are defined starting on
page 115.
Содержание 5071A
Страница 2: ......
Страница 28: ...Preface xxvi Operating and Programming Manual ...
Страница 29: ...____________________________________________ Chapter 1 Getting Started Getting Started ...
Страница 32: ...Chapter 1 Getting Started The 5071A at a Glance 4 Operating and Programming Manual ...
Страница 33: ...Chapter 1 Getting Started The 5071A at a Glance Operating and Programming Manual 5 ...
Страница 34: ...Chapter 1 Getting Started The 5071A at a Glance 6 Operating and Programming Manual 48V OPTION REAR PANEL VIEW ...
Страница 40: ...Chapter 1 Getting Started Performing Basic Tasks 12 Operating and Programming Manual ...
Страница 41: ...____________________________________________ Chapter 2 Using the 5071A Using the 5071A ...
Страница 57: ...____________________________________________ Chapter 3 In Case of Difficulty In Case of Difficulty ...
Страница 97: ...____________________________________________ Chapter 5 Command Reference Command Reference ...
Страница 119: ...Chapter 5 Command Reference Description of SCPI Commands Operating and Programming Manual 91 ...
Страница 136: ...Chapter 5 Command Reference Error Messages Codes 108 Operating and Programming Manual ...
Страница 137: ...____________________________________________ Chapter 6 Specifications Specifications ...
Страница 145: ...____________________________________________ Chapter 7 Installation Installation ...
Страница 162: ...Chapter 7 Installation Cesium Beam Tube Replacement 134 Operating and Programming Manual ...
Страница 163: ...____________________________________________ Appendix A Performance Tests Performance Tests Verifying Specifications ...
Страница 186: ...Appendix A Performance Tests 5071A PERFORMANCE TEST 158 Operating and Programming Manual ...
Страница 187: ...____________________________________________ Appendix B Glossary Glossary ...
Страница 190: ...Appendix B Glossary 5071A Glossary 162 Operating and Programming Manual ...