Commissioning
Page 27/40
Installation and starting instruction manual
SYCLOPE TRACE’O
®
2)
Using « trace » mode
The particular function of the
SYCLOPE TRACE’O
®
controller is to measure traces of chemical
products. It can be use in dialysis system to measure « no chlorine » or in a swimming pool ozone
treatment to ensure no ozone returns into the swimming pool.
The installation of the
SYCLOPE TRACE’O
®
equipments must be realized with a by-pass between the
”measurement circuit” and the “polarisation circuit” of the measurement probe.
Mysyclope.com
Internet site
Mysyclope.com / PDA
(Option)
Water to be controlled is sampled from the measurement circuit
Water for checking the sensor is sampled from the polarization circuit
Measuring cell receives the sensor for measuring chemical product and sends the value to the
controller.
Beginning of the polarisation cycle (Polarisation step) :
o
Solenoid valve A closed
o
Solenoid valve B opened
o
Waiting to reach the programmed threshold
Starting the polarization time (Maintain step):
o
Checking the good working of the sensor and waiting stabilization of the internal
chemistry
Starting recovering measurement (Depolarization step):
o
Solenoid valve A is opened
o
Solenoid valve B is closed
o
Checking time and waiting to reach the low programmed threshold.
Polarisation circuit
Measurement circuit
Magnetic valves for switching
water circuits
A
B