Page : 19
________________________________________________________________________________________________
HYDRO’pH
®
,
HYDR’ORP
®
,
HYDRO’pH Duo
®
and
HYDRO’pH/ORP
®
instructions
a)
Mounting procedure of pipe saddles
Step 1
Step 2
Step 3
b)
Mounting procedure for the sensors connection kit
Step 4
Step 5
Step 6
NOTE:
pH or ORP sensors can be mounted to 360 °.
However, we recommend performing this assembly to the vertical:
maintenance is easier to achieve!
Gasket
Screw
Nut
Apply the pipe saddle
either side of the tubing
on a straight section of at
least 15 cm.
Tighten the two sides at
the same time to keep the
pipe saddle perfectly
straight.
Once the pipe saddle set,
drill a hole of 16mm
properly centered
Sensor-holder
Put teflon on the « sensor-
holder » threads.
Teflon band
Mount the "sensor-
holder"on the pipe saddle.
Tighten gently and finish
with the wrench.
The « sensor holder » is
mounted, wait the end of
the installation to
assemble the sensors!