Mounting
The FLM must be mounted to a vertical
face or wall for function of the flowmeter
and to allow for gravity to assist in draining
the system. Mount the FLM as close to
the analyzer inlet system or analyzer as
possible.
The mounting holes are designed to fit
over the bolt heads and assist in mounting.
Attach with 1/4 in. fasteners with a
maximum head size of 1/2 in. (not provided)
in the four mounting holes on the plate.
The FLM is designed to flow with fast loop
(process) connections on the left and the
outputs to the analyzer on the right. System
operation is not recommended in any other
orientation.
Dimensions, in inches (centimeters), are for
reference only and are subject to change.
Top
7.50
(19.0)
1.50
(3.81)
13.0
(33.0)
13.5
(34.3)
4 mounting holes for
1/4 in./6 mm socket
head cap screw or
hex head bolt
0.35 (0.89)
0.40 (1.02)
1/4 in. Fast Loop Module
Left Side
Right Side
Fast loop inlet
Fast loop outlet
Relief outlet
(option)
7.15
(182)
1.71 (43.4)
11.9
(302)
9.90
(251)
0.69
(1.75)
0.92 (23.4)
4.56
(116) 5.24
(133)
7.97
(202)
11.0
(279) 12.8
(325)
Return from
analyzer
Out to
analyzer
Grab sample
outlet (option)
Purge in
Drain/vent
Drain/vent
Weight
22 to 27 lb (10.0 to 12.2 kg)
A Swagelok Pre-Engineered Subsystem Fast Loop Module
10
Содержание Fast Loop Module
Страница 51: ...Tornado Model 602 Maintenance Instructions A Swagelok Pre Engineered Subsystem Fast Loop Module 51...
Страница 52: ...Tornado Model 602 Maintenance Instructions A Swagelok Pre Engineered Subsystem Fast Loop Module 52...
Страница 53: ...Tornado Model 602 Maintenance Instructions A Swagelok Pre Engineered Subsystem Fast Loop Module 53...
Страница 54: ...Tornado Model 602 Maintenance Instructions A Swagelok Pre Engineered Subsystem Fast Loop Module 54...
Страница 55: ...A Swagelok Pre Engineered Subsystem Fast Loop Module 55...