![Svantek SV 271 LITE Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/svantek/sv-271-lite/sv-271-lite_user-manual_1409265011.webp)
SV 271 LITE User Manual
11
The safety fuse is located on the left side of the
connection plate.
Note: Do not remove the battery from the case! This operation must be done
only by the authorised service.
Note: Battery is not restricted for air, surface and water transport. Classified
as non-hazardous material (IATA/ICAO Special Provision A67, DOT-CFR
Title 49 parts 171-189, IMDG amendment 27).
Note: During station operation it is recommended to charge the internal
battery as often as possible; this will extend battery life. It is necessary to
charge the battery after any total discharge
.
Note: The station should not be stored for a long time with a discharged
battery. Storing the station with a discharged battery can damage the
battery. Storing the station with a discharged battery for more than one
month may result in impossibility to recharge the battery.
Note: If the station is planned to be stored for a long period of time, it is
recommended to charge its battery up to 100% of its capacity. The battery
should be charged at least every 6 months.
2.1.2. SD 270(A) LITE Controller
The SV 271 LITE monitoring station is equipped
with the SD 270 LITE (and later modification
SD 270A LITE) controller that integrates and
controls all system modules and is responsible for
the powering of all elements of the monitoring
station.
A very important task of the controller is power distribution - it provides appropriate DC
power to the SVAN 971, managing external power sources such as a power supply
SB 270 or optional solar panel or external battery.
The next key task of the controller is integration of the whole system
– examining the
state and condition of every module of the system and the immediate indication of all
problems on the control panel by means of several LEDs.
The controller also analyses the temperature condition inside the monitoring station
case. If the internal temperature is higher than 50
o
C, the controller will switch off
internal battery charging. If the temperature further increases over 65
o
C, the controller