CURIS
®
Instructions for Use
37
Error
no.
Meaning
Cause
Resolution
Err 41,
Err 42
During the power-up test,
the device recognized
actuation of the yellow
and/or blue finger switch of
a hand piece connected to
the monopolar output.
Activating a finger switch
before the power-up test
completes, defect in the
hand piece or its connection
cable, or device defect.
If not caused by actuation,
disconnect the hand piece
and turn the device off and
back on again. If the error is
still displayed, there is a
device error.
Err 45,
Err 46
During the power-up test,
the device recognized
actuation of the yellow
and/or blue finger switch of
a hand piece connected to
the bipolar output.
Activating a finger switch
before the power-up test
completes, defect in the
hand piece or its connection
cable, or device defect.
If not caused by actuation,
disconnect the hand piece
and turn the device off and
back on again. If the error is
still displayed, there is a
device error.
Err 47,
Err 48
During the power-up test,
the device recognized
actuation of the yellow
and/or blue foot switch
pedal.
Activation of the foot switch
before the power-up test
was complete, short circuit
in the foot switch or its
connection cable, or device
defect.
If not caused by actuation,
disconnect the foot switch
and turn the device off and
back on again. If the error is
still displayed, there is a
device error.
Err 51
During the power-up test,
the device detected the
actuation of a button on the
operator panel.
Attempting to actuate a
button before the power-up
test completes, or operator
panel defect.
If not caused by activation,
the operator panel of the
CURIS
®
is defective.
Err 52,
Err 152,
Err 53,
Err 153,
Err 55
Instrument recognition
calibration error.
Internal hardware error.
Turn the device off and back
on, and connect and
disconnect the instrument.
Contact the
manufacturer/specialist
dealer if the error remains.
Err 54,
Err 154,
Err 56
Instrument is not
recognized.
Contacting error; internal
hardware error.
Turn the device off and back
on, and connect and
disconnect the instrument.
Contact the
manufacturer/specialist
dealer if the error remains.
Err 89
Overheating of the RF
output stage noted.
Continuous RF activation for
an extended time outside
the specification (10s/30s).
Turn off the device and let it
cool down for a few minutes
before turning it on again. If
the error is still displayed,
there is a device error.