CURIS

®

 Instructions for Use 

 

 

 

4

 

Function and meaning of the display and control elements 

Front of the device 

 

 

Mains switch 

“Auto-RF control active” indicator 

Program memory P1 … P4 (to store device settings) 

CONTACT selection button for monopolar contact coagulation 

CUT 1 selection button for monopolar cutting 

SOFTSPRAY selection button for contactless monopolar coagulation (spray coagulation) 

CUT 2 selection button for monopolar cutting with coagulation 

The indicator lamp lights up during RF output in the MONOPOLAR operating mode. 

MACRO selection button for bipolar coagulation 

10 

BICUT 1 selection button for bipolar cutting 

11 

PRECISE selection button for bipolar coagulation with fine instruments 

12 

BICUT 2 selection button for bipolar cutting with coagulation 

13 

RaVoR™ selection button for bipolar RF volume reduction, for instance in snore therapy 

14 

EXCISE selection button for bipolar cutting with ring forceps 

15 

The indicator lamp lights up during RF output in the BIPOLAR operating mode. 

16 

Display of the RF cutting output setting 
(operating modes CUT1, CUT2, BICUT1, BICUT2, EXCISE) 

17 

Rotary knob to select the RF cutting output 
(operating modes CUT1, CUT2, BICUT1, BICUT2, EXCISE) 

18 

Display of the RF coagulation output setting 

Содержание CURIS 36 0 00-01

Страница 1: ...INSTRUCTIONS FOR USE CURIS 4 MHz Radiofrequency Generator REF 36 0 00 01 These instructions for use apply for all devices according to chapter 5 1...

Страница 2: ......

Страница 3: ...f 9 5 5 Self test 10 5 6 Connecting accessories 10 5 6 1 Connecting the neutral electrode 10 5 6 2 Connecting monopolar handles for cutting and coagulation 10 5 6 3 Connecting bipolar accessories 11 5...

Страница 4: ...ication 24 9 2 Recurring safety related inspections 25 9 3 Diagrams 26 9 3 1 RF output 26 9 3 2 RF voltage 28 9 3 3 Control characteristic curve 30 9 4 Guidelines and manufacturer declaration for elec...

Страница 5: ...ct current Display Indicator panel on the generator Err Error F Floating hPa Hectopascal Hz Hertz kHz Kilohertz mA Milliampere MDD 93 42 EEC EU Medical Device Directive MHz Megahertz MPG Medical Devic...

Страница 6: ...e electrodes for monitoring may not be used Do not use the active surgery electrode in the vicinity of the ECG electrodes minimum distance 15 cm Compliance with the following general rules is required...

Страница 7: ...ss only works moderately or not at all This method is used to sever or resect tissue using blade or needle shaped electrodes or wire or tape loops Electrosurgical coagulation Two active principles are...

Страница 8: ...exist in other specialist fields The CURIS is suitable for electrosurgical cutting and coagulating Exclusions are application immersed in liquids and applications that require a higher RF output than...

Страница 9: ...or a Sutter service center If this is not available packaging that properly protects the device being returned is mandatory In case of improper packaging liability rests exclusively with the sender Th...

Страница 10: ...POLAR operating mode 9 MACRO selection button for bipolar coagulation 10 BICUT 1 selection button for bipolar cutting 11 PRECISE selection button for bipolar coagulation with fine instruments 12 BICUT...

Страница 11: ...r coagulation operating modes 27 STANDARD special function button for working without automatic START STOP or TIMER functions 28 Connection socket for monopolar instruments 29 TIMER special function b...

Страница 12: ...n for electric potential equalization 47 T V seal of approval 48 Hardware and software version four characters 49 Volume control NOTE In the chapters that follow the numbers in brackets for example X...

Страница 13: ...tion 46 The required connecting cable is not included in the scope of delivery and can be obtained from us if needed 5 3 Mains supply WARNING To exclude the risk of electric shock this device may only...

Страница 14: ...o the connection socket on the front of the device 30 One piece neutral electrodes or neutral electrodes with two partial areas that make it possible to monitor the contact to the patient may be conne...

Страница 15: ...to the socket 38 on the back of the device see illustration below left The foot switch or for the coagulation electrodes the AUTO START function 26 is subsequently used for activation see chapter 6 1...

Страница 16: ...urface coagulation with sparks fulguration BIPOLAR for bipolar electrodes Cutting yellow selection buttons BICUT 1 10 Smooth cut with no coagulation zone BICUT 2 12 Cutting current with coagulation zo...

Страница 17: ...m value For the maximum values of the various operating modes see the technical information in chapter 9 NOTE Always select the lowest output setting for the desired effect On the other hand note that...

Страница 18: ...in flawless condition The surface of the device may get very hot during the extended application of RF current 6 1 Special functions The device is equipped with the following special functions The Au...

Страница 19: ...ve instrument cleaning If one switches to a monopolar operating mode during treatment also in case of accidental monopolar activation and an electrical contact is detected between the two bipolar cond...

Страница 20: ...T button 26 This step protects against unintentional activation NOTE A manually selected output setting in a bipolar cutting operating mode cannot be saved for safety reasons In this case the output s...

Страница 21: ...e chosen current with the finger switch on the electrode handle or the foot switch The operating mode indicator lamps for MONOPOLAR 8 and BIPOLAR 15 have to light up according to the current type that...

Страница 22: ...rical equipment supplied by the mains network The patient must not have any contact with grounded metal components otherwise there is a risk of point burns at the contact points Especially the patient...

Страница 23: ...ossible with a one piece neutral electrode No warning signal is triggered in case of poor contact Applying the neutral electrode to implants and other metal components or to bony protrusions and scar...

Страница 24: ...ARNING Radio equipment mobile telephones or other transmitters operated in the immediate vicinity of the RF generator may impair its safe functioning See chapter 9 4 for the minimum distances from tra...

Страница 25: ...be the scope of functionality in easily comprehensible terms The accessory has suitable plug connections that can be connected to the device without excessive force and problems The manufacturer is wi...

Страница 26: ...cutting operating mode at 0 is recommended if using this operating mode is not intended 7 9 Risks due to high electric voltage The application of a current type with a high voltage in particular a mon...

Страница 27: ...at disinfectant residues are fully removed before using the device NOTE Always keep accessories for RF devices in flawless functional condition Non functional damaged or defective accessories may pose...

Страница 28: ...0 W 20 on 300 1 20 A 4 00 MHz BICUT 2 cutting max 80 W 20 on 300 1 30 A 4 00 MHz 33 kHz EXCISE cutting max 80 W 20 on 300 1 20 A 4 00 MHz 33 kHz MACRO coag max 80 W 20 on 50 2 20 A 4 00 MHz PRECISE co...

Страница 29: ...he functionality of the rotary encoders 9 Verifying the functionality of activation the activation lamps and the activation tone 10 Verifying the functionality of the two pedal foot switch 11 Verifyin...

Страница 30: ...s authorized by us with the maintenance repair of the CURIS The service manual also lists all assemblies that are available as replacement parts 9 3 Diagrams 9 3 1 RF output MONOPOLAR Curve 1 Output s...

Страница 31: ...rve 1 Output setting maximum output Curve 2 Output setting half output RF power watts R BICUT 1 1 2 R RF power watts BICUT 2 1 2 R RF power watts EXCISE 1 2 R RF power watts MACRO 1 2 RF power watts R...

Страница 32: ...RIS Instructions for Use 28 9 3 2 RF voltage MONOPOLAR CUT 1 Display watts Peak voltage V CUT 2 Display watts Peak voltage V CONTACT Display watts Peak voltage V SOFTSPRA Y Display watts Peak voltage...

Страница 33: ...29 BIPOLAR BICUT 1 Display watts Peak voltage V BICUT 2 Display watts Peak voltage V EXCISE Display watts Peak voltage V MACRO Display watts Peak voltage V PRECISE Display watts Peak voltage V RaVoR D...

Страница 34: ...on 600 1 CONTACT on 400 2 SOFTSPRAY on 600 1 BICUT 1 BICUT 2 EXCISE on 300 1 MACRO on 50 2 PRECISE on 50 3 RaVoR on 50 Display watts RF power watts MONOPOLAR CUT Display watts RF power watts MONOPOLAR...

Страница 35: ...tromagnetic environment guidelines HF emissions according to CISPR 11 Group 2 The CURIS has to emit electromagnetic energy for its intended function Interference with devices in the vicinity is possib...

Страница 36: ...ifferential mode voltage 2 kV common mode voltage 1 kV differential mode voltage 2 kV common mode voltage The quality of the supply voltage should correspond to a normal commercial or hospital environ...

Страница 37: ...compliance levelb according to on site testing b Interference may occur in the vicinity of equipment marked with the following symbol Radiated HF interference according to IEC 61000 4 3 3 V m 80 MHz t...

Страница 38: ...ona 217 Hz 9 0 3 800 960 810 870 930 Pulse modulationa 18 Hz 28 0 3 1700 1990 1720 1845 1970 Pulse modulationa 217 Hz 28 0 3 2400 2570 2450 Pulse modulationa 217 Hz 28 0 3 5100 5800 5240 5500 5785 Pul...

Страница 39: ...e shall only be disposed of in household garbage or as hazardous waste after careful cleaning disinfection and if applicable sterilization Infected sharp components of disposable instruments are treat...

Страница 40: ...continues flashing after it is acknowledged Disconnect and check the connected accessory Accessory is in contact with tissue or defective Even though the accessory was removed the button continues fla...

Страница 41: ...ot switch before the power up test was complete short circuit in the foot switch or its connection cable or device defect If not caused by actuation disconnect the foot switch and turn the device off...

Страница 42: ...ctants commonly used in hospitals NOTE For safety reasons always perform a functional test prior to surgical application Always activate the pedals with the electrosurgery device turned on To prevent...

Страница 43: ...esigned especially for the CURIS features numerous extras such as storage baskets and a foot switch adapter It is fully assembled on delivery The CURIS is secured against sliding by engaging in the me...

Страница 44: ...CURIS Instructions for Use 40 Notes...

Страница 45: ...CURIS Instructions for Use 41 Notes...

Страница 46: ......

Страница 47: ......

Страница 48: ...Distributed by Manufacturer Sutter Medizintechnik GmbH Tullastr 87 79108 Freiburg Germany Telephone 49 0 761 51 551 0 Fax 49 0 761 51 551 30 E mail info sutter med de www sutter med de Changes reserv...