38
LAMBDA SC OPERATION MANUAL – REV. 1.14C (20190606)
Byte
Num.
Byte Value
(Decimal, hexadecimal, & binary)
Upper &
Lower
Nibble
Values
(Decimal,
hexadecimal,
& binary)
Description
5
th
0 – 9, 16 – 25, 32 – 41, 48 – 57, 64 – 73, 80 – 89,
96 – 105, 112 – 121, 128 – 137, 144 – 153.
00 – 09, 10 – 19, 20 – 29, 30 – 39, 40 – 49, 50 –
59, 60 – 69, 70 – 79, 80 – 89, 90 - 99
0000 0000 – 0000 1001, 0001 0000 – 0001 1001
0010 0000 – 0010 1001, 0011 0000 – 0011 1001
0100 0000 – 0100 1001, 0101 0000 – 0101 1001
0110 0000 – 0110 1001, 0111 0000 – 0111 1001
1000 0000 – 1000 1001, 1001 0000 – 1001 1001
0 – 9
0 – 9
0000 - 1001
Milliseconds: 100s digit (0
through 9)
0 – 9
0 – 9
0000 - 1001
Milliseconds: 10s digit (0
through 9)
6
th
0 – 9, 16 – 25, 32 – 41, 48 – 57, 64 – 73, 80 – 89,
96 – 105, 112 – 121, 128 – 137, 144 – 153.
00 – 09, 10 – 19, 20 – 29, 30 – 39, 40 – 49, 50 –
59, 60 – 69, 70 – 79, 80 – 89, 90 - 99
0000 0000 – 0000 1001, 0001 0000 – 0001 1001
0010 0000 – 0010 1001, 0011 0000 – 0011 1001
0100 0000 – 0100 1001, 0101 0000 – 0101 1001
0110 0000 – 0110 1001, 0111 0000 – 0111 1001
1000 0000 – 1000 1001, 1001 0000 – 1001 1001
0 – 9
0 – 9
0000 - 1001
Milliseconds: 1s digit (0
through 9)
0 – 9
0 – 9
0000 - 1001
Milliseconds: 0.1s digit (0
through 9)
* NOTE: If hours are set to 5, then all other time fields must be set to zero. In other words, 5
hrs 0 min. 0 sec. 0.0 ms is the maximum time for the timer commands (or 4 hrs., 59 min., 59
sec., 999.9 ms PLUS 0.1 ms).
10.4.8
TTL IN Pulse Trigger Disabled
Disables the use of TTL IN for controlling the state of the shutter.
10.4.9
TTL IN High Triggers SmartShutter to Open
When the signal coming in on TTL IN goes high, the shutter opens. Once the signal goes low
again, the shutter closes (the low signal keeps the shutter closed, and the high signal keeps
the shutter open).
10.4.10
TTL IN Low Triggers SmartShutter to Open
When the signal coming in on TTL IN goes low, the shutter opens. Once the signal goes high
again, the shutter closes (the high signal keeps the shutter closed, and the low signal keeps
the shutter open).
10.4.11
TTL IN Rising Edge Triggers SmartShutter to Toggle
On the rising edge of the signal coming in on TTL IN, the shutter opens if closed or closes if
open (toggles its state).
Содержание SmartShutter Lambda SC
Страница 3: ......
Страница 4: ......
Страница 10: ...2 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Страница 12: ...4 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Страница 14: ...6 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Страница 16: ...8 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Страница 18: ...10 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Страница 22: ...14 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Страница 30: ...22 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Страница 32: ...24 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Страница 66: ...58 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Страница 70: ...62 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Страница 72: ...64 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Страница 74: ...66 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Страница 76: ...68 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...
Страница 88: ...80 LAMBDA SC OPERATION MANUAL REV 1 14C 20190606 This page intentionally blank...