WARNINGS
AND
SAFETY
HAZARDS
OP CIC1200 E Rev4.0 06-09 Chapter 1
3
1
1.5. Emergency Off and Safe-
ty Interlocks
SUSS lamp houses and machinery (e.g. mask align-
ers) are equipped with safety related interlocks ac-
cording to there inherent risks.
Such interlocks may consist of door locks, door
switch, flow monitoring device and EMO functions.
These interlocks can switch off the CIC and / or con-
trol the CIC status using signal or power line connec-
tions to the CIC. The EMO of the host machine shall
switch of the power line to the CIC via contactor or
safety relay.
A sample of a safety circuit is shown below. For fur-
ther details refer to the manuals of the host machine
or contact SUSS service.
Lamp House Door Interlock
Sample from SUSS MA/BA8:
Opening the lamp house door will immediately open
the door switch, effectively turning off the CIC power
supply the same way as by activation of the EMO
button. The EMO-relay K6.3 will switch off the CIC
main power.
Warning!
Never defeat safety interlocks or EMO
between
the CIC and the lamp house or ma-
chine!
L1
power supply
24VDC 10Amp
for detail information
see schematic diagram
sheet 6
L1
+
-
-T2.1
10A
-F2.1
2AT
-F5.1
CIC
A1
S11
S12
S34
S33
EMO Relay
-K6.3
21
22
+11
-S6.1
13
14
Interlock connector CIC
N
PE
CIC 1200
23
24
-K6.3
33
34
1
2
3
4
5
6
-K5.1
1
2
3
4
5
6
-K5.2
Содержание CIC1200
Страница 1: ...USER MANUAL Constant Intensity Controller CIC1200 Rev4 0 06 09 www suss com...
Страница 2: ......
Страница 4: ...DISCLAIMER AND PUBLISHING DETAILS II DISCLAIMER i...
Страница 16: ...PROGRAM PARAMETERS 10 OP CIC1200E Rev4 0 06 09 Chapter 3 3...
Страница 36: ...OPERATING PROCEDURES 28 OP CIC1200E Rev4 0 06 09 Chapter 4 4...
Страница 40: ...MAINTENANCE AND SERVICE 32 OP CIC1200E Rev4 0 06 09 Chapter 5 5...
Страница 41: ......