Using and Controlling the AirSeal® iFS in its Different Modes
23
EN
• Blunt obturator with suture anchor
• cannula and Low Profile Obturator with Bladeless Optical Tip
The AirSeal® Optical Obturator and Cannula System is composed of a sterile sin-
gle patient use instrument consisting of an obturator
(1)
and a cannula
(2)
(see
Fig. 6-1, Fig. 6-2, and Fig. 6-3). The obturator may be used with or without visual-
ization for primary and secondary insertions.
If the Optical cannula is being used for primary entry, use of a laparoscope to pro-
vide visualization to enhance safe abdominal entry is advised.
1. Use sterile techniques to remove the instrument from the package.
2. The obturator
(1)
and the cannula
(2)
(see Fig. 6-1, Fig. 6-2, and Fig. 6-3) are
packaged together.
3.
When using optical obturator
• Connect an appropiate 0° endoscope to the light supply and monitor (see
manufacturer's instructions).
• Insert the endoscope into the opening at the proximal end of the obturator
until it reaches the distal tip of the obturator.
6.1.2
Initial Insufflation
When working in the
AirSeal Mode
, it is possible to initiate insufflation either
through the open cannula, through the cannula with inserted obturator or
through a Veress needle.
CAUTION!
The manufacturer strongly advises to begin the insufflation by using the Veress
needle for intra-abdominal procedures. For trans-anal procedures, insufflation
should be performed through a standard Luer port or AirSeal® Access Port.
Initial insufflation with Veress needle entry or AirSeal® Cannula
1. Switch device on (consult section 5.5 "Switch Device On" for additional infor-
mation).
Fig. 6-2
AirSeal® Blunt Obturator with
Suture Tie Down
(1)
Blunt obturator
(2)
Cannula
(3)
Manifold plug
(4)
Spring anchor
(3)
(1)
(4)
(2)
Fig. 6-3
AirSeal® Cannula and Low Pro-
file Obturator with Bladeless
Optical Tip
(1)
Bladeless optical tip obturator
(2)
Cannula
(3)
Manifold plug
(4)
Cannula latch
(3)
(1)
(4)
(2)
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...