RMK-700 Operating Manual
2
Safety symbol and instructions
Note: Twist screws here before installation
Working condition :
1
、
No hard light shoot directly
2
、
No electromagnetic field of strong interference
3
、
Temperature : +5
℃
~ +40
℃
4
、
Relative Humidity : 30% ~ 75% RH
5
、
Atmospheric pressure range :
700hpa~1060hpa
Working environment
1
、
Do not hit or drop the instrument. The impact may cause damage to the function of this
instrument. Please handle with care.
2
、
An exposure to the direct sunlight or bright indoor lights may influence the result of the
measurement. Please use the URK-800 in an appropriate optometry room.
3
、
Please contact your dealer while connecting this to other associated equipment.
4
、
A sudden heating will cause condensation on the protective glass in the monitor screen
and on optical parts inside the instrument. In this case, wait until condensation
disappears before performing measurements.
5
、
Keep the objective glass on the examinee side clean. If smudged, it may cause an
ERROR or inaccurate measurements.
6
、
In case there is smoke, strange odor or noise during operation, disconnect the power
supply and consult the distributor.
7
、
Don’t use organic solvents such as alcohol, paint thinner, benzene, etc. to clean the
surface of this instrument. It may damage the instrument.
8
、
When moving the RMK-700, Fix the stage by using stage holding knob and clamping
bolt, disconnect the power cable, and then lift the bottom of the unit with both hands.
9
、
When the system is not in use for a longer period of time, disconnect the power supply
and cover with the dust cover.
10
、
Some material using instructions touched directly with skin
:
when operating the
instrument,
the
customers
need
to
use
medical
non-woven
fabrics(Specification:8cm*8cm)to separate the touching part between instruments and
patients, to avoid patients touch directly instrument surface.
11
、
Don’t open the enclosure until our company agree.