RMK-700 Operating Manual
21
4.1 As the power switch is on
Message
Cause
Method of settlement
Motor Error
Internal
abnormality
for the
equipment
Re-input the power in 10
seconds after switching
it off. In case that the
message is indicated
again, contact our sales
representative.
EEPROM Error
EEPROM Data Error
System Error
Clock Error
INVALID SETUP DATA
– REF
Abnormality in
the internal data
for
Refractometry
Please contact our sales
representative.
4.2 Messages during measurement
Message
Cause
Method of
Settlement
TRY
AGAIN
Refer the 10th page
Refer the 10th
page
Objective glass in the measurement
window is polluted
Clean the glass
+ OUT
Sphere of examinee’s eye exceeds
+15m
-1
Impossible
to
measure
Object lens within measurement
window is polluted
Clean the glass
- OUT
Sphere of examinee’s eye exceeds -15
m
-1
Impossible
to
measure
Object lens within measurement
window is polluted
Clean the glass
C OUT
Sphere of examinee’s eye exceeds 6 m
-1
Impossible
to
measure
Object lens within measurement
window is polluted
Clean the glass