6.4 F2 Calibration
6.4.1 Static Calibration
Operation Steps:
Step0: Press the button
【
#1
】
/
【
#2
】
/
【
#3
】
/
【
#4
】
to let its status bar being red for selecting ‘Weigher
No.’.
Step1: Zero Calibration. Let the weigher at unloading and static state, after the real-time weight display
value being stable, press the button
【
Zero Cal.
】
to display and save the new
Zero Value.
If the new Zero
Value is not in its allowed range, the operation
【
Zero Cal.
】
will be invalid.
Step2: Load Calibration. Load standard weights on the weigher with ‘Loading weight
≥
[50% × Scale
Capacity]’, after the real-time weight display value being stable, press the button
【
Load Cal.
】
to input
‘Calibrating Weight’ value, then the new
Span Coefficient value will be displayed and saved.
If the loading
weight is too small [AD Value
≤
Zero Value], the operation
【
Load Cal.
】
will be invalid.
Sign
Data
Sign
Data
Big Digits
Real-time Weight [Weight Unit].
Zero Value
The saved ‘Zero Value’.
Span Coeff
The saved ‘Span Coefficient’ value.
Zero Var
Zero Variation = AD Value – Zero
Value.
Cal. Weight
Calibrating Weight = Weight of
Standard Weights
AD Value
Real-time AD Value of weighing
signal.
Motion
Stable
Weight is dynamic changing.
Weight is stable.