Note:
Total Capacity of Loadcells = Loadcell Capacity × Loadcell Number.
Voltage Ratio = Excitation Voltage on the terminal of controller / Excitation Voltage on the terminal of
loadcells.
The rated excitation voltage for loadcells is DC10V. It’s best to measure the actual voltage value.
4-wire connection: The voltage attenuation is big, the voltage on both sides should be measured.
6-wire connection: The voltage attenuation is small, Voltage Ratio can be set to 1.0000.
RUN
NUM
PCS
TOT
SET
GROSS
NET
SUM
LOCK
SUPMETER
ZERO
NZ
SP1
SP2
SP3
FILL COMP
DEV FALL PACK
HH
HI
LO
LL
0
DISC
【
MENU
】
: Exit
【
ENT
】
: Save
M
【
▲
】
【
▼
】
:
105: New Span Coefficient value (Max. Display Value: 99.9999).
oLd: Original Span Coefficient value.
Sr: Span Correction Ratio = New Value / Original Value (Display
Range: 0.00001~9.99999).