25
6.3 F1-SET Parameter Setting
6.3.1 Weighing Parameters (SCAL)
No.
Sign
Range
Default
Description
Set
100
UnIt
0~3
2
Weight Unit
0: None
1: kg
2: t
3: g
101
dot
0~4
ooo.oo
Decimal Point
0: ooooo
1: oooo.o
2: ooo.oo
3: oo.ooo
4: o.oooo
102
SCALE
1~999999
2000
Scale Capacity
Max. loading weight of the load receptor.
Scale Capacity
≤
(Loadcell Capacity × Loadcell
Quantity) – Load Receptor Weight.
When ‘Weight value = Scale Capacity’, AO of
weight signal will output current 20mA.
103
dIV
1~500
1
Scale Division
1, 2, 5, 10, 20, 50, 100, 200, 500
104
ZEro
-20000~
+999999
0
[*]
Zero Value
Only for query.
105
SPAn
>0
1.0000
[*]
Span Coefficient
Max. display value: 99.9999.
Only for query.
106
Stb.r
0~500
1
Stablity Judging Range
[Division]
Set value = 0: No ‘Stablity Judging’.
Set value > 0: Weight Variance per [107] ‘Stablity
Judging Time’ being in [106] ‘Stablity Judging
Range’ means ‘Weight is stable’.
107
Stb.t
0.5~5.0
1.0
Stablity Judging Time
[s]
[*]: ‘RAM Reset’ operation has no effect on the parameter.