Chapter 7 Configuration Settings
www.supmea.com
20
7.2.5 Dissolved oxygen signal setting
7.2.5.1 Online calibration:
Zero oxygen and air calibration can be selected. The first point of the calibration
point corresponds to zero oxygen calibration, which can be calibrated in
anhydrous sodium sulfite solution (configured with deionized water), and the
second point corresponds to air calibration, which can be in air or saturated air and
water To calibrate, after entering the calibration interface, you need to wait for a
while, and after the measured value is stable, use to complete the calibration.
Figure 31
7.2.5.2 Signal Revision
The measured dissolved oxygen value can be corrected.
Figure 32
--Signal Setting--
1.Online Calibration
2.Signal Correction
3.Salinity Setting
--Signal Correction--