Technical Specifications
●
Measuring range (Moisture): 5 to 50% Depending on crop type
●
Operating range (Temperature): 0 to 50 ºC
●
Resolution: 0.1% moisture
●
Temperature compensation: Automatic
●
Weight including case: 2 kg
●
Dimension: 13 x 21 x 8 cm
●
Battery: 9 V Alkaline
●
Display: 2 line LCD with backlight
●
Accuracy: Better than +/- 0.25% in normal moisture range, by using average and
calibration adjustments
Troubleshooting
➢
The meter does not turn on by pressing the ON/TEST button
Change the battery. It is a good idea to always have a new battery in your carrying-case
➢
The meter shows Lo<xx% in the display
The moisture content of the sample in the measuring cell is below the measuring range
➢
The meter shows Hi>xx% in the display
The moisture content of the sample in the measuring cell is over the measuring range
Or
The two grinding discs are touching one another and there must be carried out a new measurement
with an increased sample. Try with approx. 2-3 ml more than the quantity specified in the display. In
particular when measuring maise or grass seed 18ml= 2x9ml must be measured out for grinding as
stated in the display.
➢
The meter shows other values compared to official certified meters
Calibrate the tester as explained in the section “Adjusting the calibration of the meter”
Ver 1.1.0
GB